- 22 -
Step 4
Fully open both the INLET and OUTLET
valves.
Step 5
While holding the ceramic ball (to ensure that
the ball does not fall out), tilt the tester so
that the output connection is pointing down-
ward.
Step 6
Pressure the ball into its socket to provide an
output pressure sufficient to purge the sys-
tem.
Step 7
After completing Steps 1 - 6, disconnect the
air supply from the INLET and connect the air
supply to the OUTLET. DO NOT APPLY PRES-
SURE AT THIS TIME.
Step 8
Using a screw driver, remove the ceramic
ball and restriction tube. The restriction tube
is located under the ceramic ball.