radEAGLET-R
User Manual
Powering the
radEAGLET-R
To remove the
powerCELL
…
1. Unscrew both locks of the cover by turning them counter-clockwise. You can use the
attached bracket or, for example, a coin.
You don’t need to remove the screws completely, just turn them until the cover
loosens.
2. Remove the cover (Fig.
, p.
Locking Screws
powerCELL with Flap
Figure 130:
powerCELL
in compartment with cover removed
3. Grab the flap and pull the
powerCELL
out of the instrument.
To install a
powerCELL
…
1. Slide the
powerCELL
into the compartment, contact side ahead.
The case of the
powerCELL
is asymmetric. It fits into the compartment in the
correct orientation only.
2. Put the cover on the compartment.
3. Lock the cover by turning both locking screws clockwise.
8.3 External Power Sources
You can connect the
radEAGLET-R
to external power sources via the connector on the rear
side of the instrument Fig.
, p.
Included with the
radEAGLET-R
are:
• Wall power supply (
, p.
)
• Car adaptor
© innoRIID GmbH • 2020-11-18
Software 3.2.12r • Document 3.3.3b4o
123/163
Summary of Contents for Ortec radEAGLET-R 2SG
Page 1: ...radEAGLET R User Manual Software 3 2 12r Document 3 3 3b4o 2020 11 18...
Page 12: ...12 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 22: ...22 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 42: ...42 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 94: ...94 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 120: ...120 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 144: ...144 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...