8.2 Replacing Batteries of the
radEAGLET-R
radEAGLET-R
User Manual
Contacts
Insertion Direction
Figure 128:
The standard black rechargeable
powerCELL
of the
radEAGLET-R
.
Rechargeable batteries should not be completely discharged. If you don’t use them
for a while, make sure to charge it periodically as specified in Appendix
, p.
Do not unscrew the black
powerCELL
. There are no user servicable parts inside.
To check the status of a black
powerCELL
…
1. Press the marked button on the contact side of the black
powerCELL
(Fig.
p.
2. Observe the LED. It flashes in a pattern proportional to the current status:
• No light: Empty
• Continuous light: Full
Press
<25 %
50 %
75 %
100 %
Contacts
LED
Figure 129:
Status check of the black
powerCELL
of the
radEAGLET-R
.
8.2 Replacing Batteries of the
radEAGLET-R
You can replace the
powerCELL
of the
radEAGLET-R
.
You can replace
powerCELL
while the
radEAGLET-R
is running. An internal buffer
provides power for at most 30 minutes. It is recharged automatically.
122/163
Software 3.2.12r • Document 3.3.3b4o
© innoRIID GmbH • 2020-11-18
Summary of Contents for Ortec radEAGLET-R 2SG
Page 1: ...radEAGLET R User Manual Software 3 2 12r Document 3 3 3b4o 2020 11 18...
Page 12: ...12 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 22: ...22 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 42: ...42 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 94: ...94 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 120: ...120 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...
Page 144: ...144 163 Software 3 2 12r Document 3 3 3b4o innoRIID GmbH 2020 11 18...