Ametek ORTEC MOBIUS-B User Manual Download Page 8

PROPERTY OF AMETEK, INC. 

 ORTEC

®

 

ORTEC

 

®

 

MÖBIUS™ Recycler HPGe Detector Cooling System 

805709F / 1115 

2

 

The unit can be connected to an uninterruptable power supply (UPS) to sustain operation during short-duration power 
failures. Fully cooled, it draws about 150 watts of power from a UPS without a “smart” capability. Alternatively, the 
MÖBIUS can run off of a “smart” UPS with a dry contact relay interface, such as the APC SMT750 UPS equipped with 
their optional AP9613 dry contact management card. In this configuration, if power fails the UPS signals the MÖBIUS 
to enter its power-saving mode to conserve UPS battery life. The cryocooler turns off, reducing power consumption to 
less  than  10 watts,  while  the  remote  status  display  continues  to  provide  all  status  readouts  including  the  liquid 
nitrogen level in the dewar. 

 

1.3.

 

SPECIFICATIONS 

1.3.1.

 

Performance 

● 

Refill Period

   After the detector is cool and the LN2 is topped off, the MÖBIUS system can run for 2 years or 

longer  without  filling.  This  is  based  on  the  sealing  technology  and  assumes  system  operation  in 
uninterrupted, steady-state conditions. 

● 

Maintenance

   Wash or replace inlet air filter every 3 months (replacements available from ORTEC). No other 

maintenance is required. 

● 

Remote Status Display

    Readout  unit with hardwired umbilical cord can be remotely  mounted to improve 

visibility  when  the  detector  is  placed  in  a  lead  shield.  The  display  registers  the  liquid  nitrogen  level  in  the 
dewar,  power  level  being  supplied  to  the  cooler,  pressure  level  within  dewar  and  general  status  of  the 
system.  

● 

Internal Controller

   Control electronics, power  supply, and cooler electrically isolated from the dewar and 

protected from LN2 spillage. 

● 

LN2  Level  Sensing

      Capacitance  level  sensor  provides  continuous  measurement  of  the  LN2  level  from           

0–100% full. 

● 

Use with UPS

   The MÖBIUS has a small onboard computer that can receive an input, via the rear-panel 

I/O

 

connector, in the event of power loss. The input causes the system to enter power saving mode and turn off 
the cryocooler. With the cryocooler turned off, the onboard computer can be powered by a small UPS. 

● 

LN2  Loss  Rate

      Approximately  3 L/day  with  non-PopTop  cryostats  when  cooler  is  off  (same  as  for 

conventional  dewar);  loss  rate  will  vary  depending  on  ambient  conditions  and  detector  size.  With  PopTop 
cryostat typical consumption can be as high as 4L/day when cooler is off.  

● 

Spectroscopic Performance

 

 

For  new  ORTEC  detectors  purchased  with  the  MÖBIUS:  at  energies  above  15 keV  we  certify  no 
degradation of the detector  resolution warranted on the detector  data  sheet, and below 15 keV, less 
than 10% of the warranted resolution.  

 

For existing ORTEC detectors in dipstick cryostats, where the MÖBIUS replaces an existing dewar or X-
Cooler®, resolution degradation is based on the existing detector performance. It is expected that the 
detector  will  have  similar  performance  compared  to  using  a  standard  LN2  dewar.  Some  degradation 
may  occur  depending  on  the  detector's  age  and  configuration.  Note:  Detector  in  a  PopTop  cryostat 
requires a 1.50 inch in diameter dipstick when converting from X-Cooler® to MÖBIUS.  

 

MÖBIUS can be used with existing non-ORTEC detectors in dipstick  cryostats having 1.25 or 1.50 inch 
diameter. It is expected that the detector will have similar performance compared to using a standard 
LN2 dewar. Some degradation may occur depending on the detector's age and configuration. 

 

Summary of Contents for ORTEC MOBIUS-B

Page 1: ...ORTEC M BIUS Recycler Condensing Liquid Nitrogen HPGe Detector Cooling System User Manual Printed in U S A ORTEC Part No 805709 0514 Manual Revision F...

Page 2: ...482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause...

Page 3: ...N 9 3 1 ASSEMBLING THE SYSTEM 10 3 2 FILLING WITH LN2 11 3 3 SAFETY AND EQUIPMENT WARNINGS 13 3 3 1 Liquid Nitrogen Safety 13 3 3 2 Other Precautions 14 3 3 2 1 HANDLING POPTOP DETECTOR CRYOSTAT ASSEM...

Page 4: ...TEC technical support personnel and authorized in writing in advance of such modifications SAFETY INSTRUCTIONS AND SYMBOLS This manual contains up to three levels of safety instructions that must be o...

Page 5: ...805709F 1115 ORTEC M BIUS Recycler HPGe Detector Cooling System v Intentionally blank...

Page 6: ...PROPERTY OF AMETEK INC ORTEC ORTEC M BIUS Recycler HPGe Detector Cooling System 805709F 1115 vi M BIUS B w HJ style cryostat M BIUS B w PopTop cryostat M BIUS w PopTop cryostat...

Page 7: ...manual 1 1 HOW M BIUS WORKS The M BIUS has a computer controlled pressure system that maintains a constant and stable 0 5 psi 3 5 kPa in the vapor space of the dewar A pressure transducer feeds the pr...

Page 8: ...pply and cooler electrically isolated from the dewar and protected from LN2 spillage LN2 Level Sensing Capacitance level sensor provides continuous measurement of the LN2 level from 0 100 full Use wit...

Page 9: ...ded for indoor use Intended for altitude up to 2000 m 6562 ft Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages up to the levels of overvoltage category II Pollu...

Page 10: ...ontains a universal ac mains power cord and a set of region specific outlet Fuse kit contains fuse block and fuses use the 250 V 10 A time delay slo blo fuses for 120 V ac mains power and the 250 V 5...

Page 11: ...805709F 1115 ORTEC M BIUS HPGe Detector Cooling System 5 Figure 3 M BIUS Top Panel Features 2 Figure 4 Rear Panel Features...

Page 12: ...e etc Replace the cover and tighten the screws just enough that they will not vibrate loose Do not over tighten This seal is intended as an air dam not a pressure seal so tighten the screws just enoug...

Page 13: ...PGe Detector Cooling System 7 Figure 6 Remove Protective Foam Block and Caution Tag Under Cooler Cover and Keep for Future Use Figure 7 Underside of Fuse Block is Keyed Figure 8 Insert Keyed Side Down...

Page 14: ...niversal power cord adapts for almost any mains power configuration Simply choose the plug that matches your outlet Figure 9 and press it onto the power cord The other end is a standard female IEC con...

Page 15: ...ight involves the interplay of shield height adjustment typically requires a hydraulic jack and wrench the thickness of acoustic damping material and or other spacers under the dewar and the insertion...

Page 16: ...the spacers damping materials M BIUS and detector cryostat assembly beside the shield to make preliminary height estimates cryostat and shield height adjustments etc Otherwise place the M BIUS under...

Page 17: ...ill show X depending on the level of LN2 currently in the dewar 4 With the remote status display clearly visible open the two fill vent valves on the front of the dewar Figure 2 This will depressurize...

Page 18: ...rbed overnight to allow the detector and M BIUS to completely cool down During cooldown leave the M BIUS fill vent valves open 11 A warm detector causes a significant amount of LN2 to boil off at firs...

Page 19: ...ther remediation unless authorized to do so by our ORTEC Global Service Center If water accumulates in the dewar remove the LN2 obtain a cylinder of high purity oil free dry nitrogen gas not house nit...

Page 20: ...r Precautions 3 3 2 1 HANDLING POPTOP DETECTOR CRYOSTAT ASSEMBLIES Grasp the assembly just beneath the base of the detector capsule where the cable exits the base plate or by the preamplifier shroud s...

Page 21: ...re re cooling Typical warm up time for most detectors is 24 hours For higher efficiency detectors 70 allow 36 hours warmup time To determine if the detector is cool enough to be operated attempt to tu...

Page 22: ...ower levels which can adversely affect detector performance due to increased vibration Therefore ORTEC recommends cleaning the filter every 3 months more often in areas with higher than normal dust le...

Page 23: ...n a cylinder of high purity oil free dry nitrogen gas not house nitrogen Connect to the fill valve and stream at very low pressure to gently boil off the LN2 and evaporate any water from inside the de...

Page 24: ...es to cool down e g 12 to 36 hours or more 4 5 REMOVING THE DETECTOR CRYOSTAT ASSEMBLY This procedure requires two people and a step stool If the detector is still cold the person working beneath the...

Page 25: ...you place it on a surface such as a bench chock it so it cannot roll If placed in contact with insulating materials such as foam the o ring in the detector capsule may freeze and breach the capsule va...

Page 26: ...PROPERTY OF AMETEK INC ORTEC ORTEC M BIUS Recycler HPGe Detector Cooling System 805709F 1115 20 Intentionally blank...

Page 27: ......

Page 28: ......

Reviews: