12
User Guide
EN
Dansensor
®
LeakPointer
H
2
O
Filling of water
General
Water temperature should be max. 50°C.
Add 1-2 ml non-alkali detergent to the water, to conserve it and make it softer.
If local water is very calcareous, we recommend to use de-mineralised water.
NOTE! Frequent cleaning is recommended to avoid bad hygiene and complicated
cleaning - see
"5. Cleaning and maintenance" on page 22
.
In general, it is always important to consider how much water should be in the tub
in
relation to the types of packages to be tested:
- If the purpose is to test relatively small packages, the tub should be filled so that the
surface is slightly above the underside of the hole-plate
when lid
is closed (approx.
25 liters - check scale
).
- If it is larger packages that can expand a lot during the test, the amount of water in the
tub must be adjusted so that the water level in the tub is somewhere between the
underside of the hole-plate
and the “Max” marking on the scale
when the package
is fully expanded during the test.
There are two ways to fill the tub with water. It can be filled either manually (see
by using the vacuum function (see
Summary of Contents for mocon Dansensor LeakPointer H2O