2. Safety precautions
Read and understand this manual for your own safety’s sake and for the quality of the work
carried out using this equipment.
The following terms are used:
"DANGER"
to designate a condition which MAY involve mortal danger of serious
injury
"CAUTION"
to designate a condition which MAY involve serious injury
"NOTE"
to designate a condition which MAY involve faulty operation of the
equipment thus causing fault measuring results
Follow the below instructions carefully:
DANGER:
Opening ISM-3 may involve mortal danger or serious injury as live
high-voltage leads may be exposed. By opening the ISM-3, the
mains supply must always be switched off. By connecting the net
ground must always be connected.
CAUTION:
If ISM-3 is moved or transported, make sure it is note exposed to
shocks or impact.
Only use water or a mild soap lather to clean the front panel.
Always make sure that there is at least 5 cm of free air behind the
ventilation holes on the back of ISM-3.
Our warranty does not cover unless the above instruction is strictly
adhered to.
NOTE:
Always install a hose on the "GAS OUT" connection and make sure
that the measuring gas is transported away from the ISM-3, as the
surrounding atmospheric air is used as reference. In order to observe
the CE`s EMC regulation all connections to the ISM-3 (not mains
connections) must be protected. The screen must end in a PG sleeve
on the back of ISM-3.
Recommendations
In order to prolong the life of the analyser it is recommended to let ISM-3 be constantly
connected to the mains voltage.
Every 6 months it is recommended to control ISM-3`s calibrations by measuring on gases
with known content of O
2
.
Recommended control - and calibration interval is 12 months.
Summary of Contents for mocon Dansensor ISM-3
Page 1: ...P N 200114 I 09 2021 UserManual...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 6: ...4 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 10: ...8 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 12: ...10 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 20: ...18 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 28: ...26 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 32: ...30 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 33: ...This blank page has been inserted to enable double sided printing of the document...
Page 34: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...