background image

Das verknüpfte Bild kann nicht angezeigt werden. Möglicherweise wurde die Datei verschoben, umbenannt oder gelöscht. Stellen Sie sicher, dass die Verknüpfung auf die korrekt  

 SPECTRO MIDEX 

µ-XRF Spectrometer 

Original operating instructions 

Summary of Contents for MID05

Page 1: ...ild kann nicht angezeigt werden M glicherweise wurde die Datei verschoben umbenannt oder gel scht Stellen Sie sicher dass die Verkn pfung auf die korrekt SPECTRO MIDEX XRF Spectrometer Original operat...

Page 2: ......

Page 3: ...erview of controls 15 5 2 Displays 16 5 3 Switching on the instrument 17 5 4 Switching off the instrument 19 5 5 Carrying out the measurement 21 5 5 1 Preparing the measuring procedure 21 5 5 2 Start...

Page 4: ...te the atoms of the sample causing them to emit radiation This radiation is measured by a semiconductor detector The instrument comes with a preset method of analysis installed Data are already stored...

Page 5: ...rrect distance The instrument is available with various stop configurations Preset stop Software controlled selectable stop Fig 1 No Component No Component 1 PC 6 Rear side 2 Instrument a Key switch w...

Page 6: ...he instrument If required additional important information is available in connection with these symbols Warning Risk of injury Warning Risk of electric shock Warning Laser beam Warning Danger of pois...

Page 7: ...epair activity The checking procedure is carried out in accordance with DGUV V3 or an equivalent national regulation Comply with the maintenance schedule specified by the manufacturer and record all m...

Page 8: ...2 5 Residual risks This instrument has been manufactured in accordance with the state of the art Nevertheless a residual risk cannot be completely excluded Information The instrument produces X ray r...

Page 9: ...Your eyes will suffer damage if you look at this laser pointer for an extended period of time Therefore never look directly at the laser pointer Information The safety circuits in the instrument are d...

Page 10: ...dated April 30 2003 last amendment dated December 11 2014 Before operating the instrument approval must be obtained from the supervisory authority in the respective country of installation if such rul...

Page 11: ...bject to impact and vibration loads Operating voltage range 90 253 V AC 50 60 Hz Spectrometer power consumption without PC Stand by mode 115 W 50 W Fuse rating 16 A slow blow Temperature range Optimum...

Page 12: ...isk of damaging the instrument Contact the SPECTRO Service Department in this matter Compliance with the following information is required regarding the instrument s transport and setup Always transpo...

Page 13: ...uding the PC is connected to the UPS Please do not connect the laser printer to the UPS Information The first installation of the UPS must be performed by a technician who is authorized by SPECTRO If...

Page 14: ...Automatic circuit breaker 6 A 3 Power connection points for instruments 6 UPS power supply connection point When changing or reconnecting the UPS proceed as follows 1 Connect all components with the U...

Page 15: ...rol element Function 1 Pushbutton Used to start a measurement 2 Height adjustment wheel Used to adjust height of scissor lift table 3 Sample chamber lid Used to open and close the sample chamber 4 Key...

Page 16: ...Operation 16 SPECTRO MIDEX 05 10 2017 5 2 Displays Fig 4 No Display Function 1 X ray radiation Lit orange when X ray tube is in operation...

Page 17: ...cal position 2 Switch on the computer The operating system is booted 3 Click the XRF Analyzer Pro button on the screen to start the routine dialog of the XRF Analyzer Pro measurement program 4 Log in...

Page 18: ...low the instructions on the screen 7 Wait until the initialization procedure for the instrument has been carried out and then click on Activate Standby Mode to switch the instrument to standby mode Th...

Page 19: ...nder standby conditions for at least 20 minutes Start measuring only after this period If you do not work with the system for an extended period of time the X ray tube must be operated in standby mode...

Page 20: ...nd then click on Close Window Alternatively you can select the Close symbol on the top right of the screen 5 Turn the key switch to the horizontal position 6 Select the Start menu in the Windows user...

Page 21: ...due to laser radiation The instrument uses a laser pointer of protection class I Your eyes will suffer damage if you look at this laser pointer for an extended period of time Therefore never look dire...

Page 22: ...g dial 4 to adjust its position until it is at the correct height 1 Fig 7 The setting procedure is indicated on the screen The correct height is reached when the beam of the laser is directed at the c...

Page 23: ...tion is confirmed by the software program in the header Information Measure a reference sample at regular intervals This will allow you to identify changes of the measurement results in good time 5 5...

Page 24: ...epairs must only be performed by service personnel authorized by the instrument manufacturer The maintenance intervals specified in the following table are guidelines which should be observed as a min...

Page 25: ...the recalibration aid in the instrument and start the energy channel recalibration procedure MCA Chapters 6 2 1 and 6 2 2 Global recalibration depending on the results of the control sample measuremen...

Page 26: ...oles 2 points towards you Fig 10 2 Fit the recalibration aid to the measurement chamber according to the three available openings Fig 11 3 When the recalibration aid has been properly fitted the laser...

Page 27: ...In the software program select the Recalibration menu item beneath the green Start button and then select MCA Recalibration The corresponding program window is displayed 3 Click on OK followed by STA...

Page 28: ...ration procedure is started manually In order to recalibrate the instrument proceed as follows 1 Ensure that the recalibration aid has been installed see Section 6 2 1 2 In the software program select...

Page 29: ...rly Insert in the power supply plug properly The monitor cannot be switched on The power plug is not plugged in properly Insert in the power supply plug properly The monitor is not on Switch on the mo...

Page 30: ...ase contact your Service Center When starting the measurement the high voltage is shut off Defective safety switches or defective tube Please contact your Service Center No spectrum is displayed when...

Page 31: ...ent The use of non genuine spare parts will make the CE Declaration of Conformity null and void Excepted are spare parts from manufacturers approved by SPECTRO 9 De commissioning Only organizations or...

Page 32: ...EC Declaration of Conformity 32 SPECTRO MIDEX 05 10 2017 11 EC Declaration of Conformity...

Reviews: