6.0     Returning the pressure calibration set 
 for 

service 

 

When returning the pressure calibration set to the manufacturer for 
service, please enclose a fully completed service information form. 
Simply copy the form on the following page and fill in the required 
information. 
The pressure calibration set should be returned in the original 
packing. 
 
Always secure a traceable certificate for confirmation of re- 
calibration. 
 
 

Summary of Contents for MGC-HIGH

Page 1: ...User Manual Portable Pressure Calibration Set MGC LOW MGC HIGH Copyright 2007 AMETEK Denmark A S...

Page 2: ...scribed and accessories may be ordered from the list of accessories Safety symbols This manual contains a number of safety symbols designed to draw your attention to instructions which must be followe...

Page 3: ...tructions 11 4 1 Operating the MGC LOW 11 4 1 1 Media to be used 13 4 1 2 After calibration 14 4 2 Operating the MGC HIGH 15 4 2 1 Media to be used 18 4 2 2 After calibration 18 4 3 Operating the opti...

Page 4: ...tion This expertise is reflected in our products which are all designed for daily use in an industrial environment Please note that we would be very interested in hearing from you if you have any idea...

Page 5: ...hnical assistance Please contact the dealer from whom you acquired the pressure calibration system if you require technical assistance Guarantee According to current terms of sale and delivery This gu...

Page 6: ...n those described in this manual and for any application other than precision pressure calibration jobs The pressure system should only be used by TRAINED PERSONNEL Never subject the pressure gauges t...

Page 7: ...product liability only applies if the pressure system is subject to a manufacturing defect This liability becomes void if the user fails to follow the maintenance instructions set out in this manual o...

Page 8: ...0 40 BAR Kg cm 0 100 BAR Kg cm and 0 300 BAR Kg cm System 3 0 10 BAR Kg cm 0 40 BAR Kg cm and 0 100 BAR Kg cm System 4 0 10 BAR Kg cm 0 40 BAR Kg cm and 0 300 BAR Kg cm Both models are supplied with...

Page 9: ...against the list shown below If any of the parts are missing or damaged please contact the dealer who sold the pressure calibration system You should receive 1 pump system T 620 incl 0 5 m hose MGC L...

Page 10: ...10 02 10 2007 124945 01 3 adapters 1 4 x 1 4 BSP 1 2 x 1 4 BSP 1 4 x 3 4 BSP MGC HIGH 1 plug 1 4 BSP 2 bonded seal 1 4 BSP 1 traceable certificate per gauge 1 user manual...

Page 11: ...INED PERSONNEL Never subject the pressure gauges to overpressure In case of overpressure the gauge should not be used for calibration purposes None of our calibration systems are cleaned or prepared f...

Page 12: ...pressure should be indicated If this is not the case the pressure gauge may have been subjected to over pressure and should not be used for calibration purposes 2 SYSTEM 1 0 6 Kg cm2 0 16 Kg cm2 0 16...

Page 13: ...be released from the system by loosening the connectors pos 5 It is now possible to pump out oil and air until no air is left in the oil 4 If other connectors than those supplied are needed please use...

Page 14: ...KYDROL or the like 4 1 2 After calibration 1 Open the pump valve pos 1 2 Make sure that the pressure system is depressurised before all items are disconnected 3 Disconnect all items 4 Close the pump v...

Page 15: ...L Never subject the pressure gauges to overpressure In case of overpressure the gauge should not be used for calibration purposes None of our calibration systems are cleaned or prepared for OXYGEN OBJ...

Page 16: ...is is not the case the pressure gauge may have been subjected to over pressure and should not be used for calibration purposes SYSTEM 1 0 6 Kg cm2 0 16 Kg cm2 0 10 Kg cm2 0 40 Kg cm2 0 10 Kg cm2 0 40...

Page 17: ...rs pos 6 It is now possible to pump out oil and air until no air is left in the oil 4 If other connectors than those supplied are needed please use fittings with correct pressure rating and correct th...

Page 18: ...1 Open the pump valve pos 2 on the right side of the pump 2 Open the pump valve pos 3 on the front of the pump gently to protect the gauges from the pressure shock 3 Make sure that the pressure system...

Page 19: ...any purposes other than those described in this manual Do not use this system for any applications other than precision pressure calibration jobs Never subject the valve system to overpressure Do not...

Page 20: ...stems with gauges are prepared for test fitted with a valve with test connector To operate this kind of system a Lloyds valve system with clamp from AMETEK Denmark A S can be used Pressure system with...

Page 21: ...lve system pos 3 5 Connect the Lloyds valve system to the valve with test connector located on the system which is to be tested pos 4 6 Open the valve on the system to be tested pos 5 7 Open the Lloyd...

Page 22: ...TEK Denmark s liability ceases if parts are replaced repaired using spare parts which are not identical to those recommended by the manufacturer non original parts are used in any way when operating t...

Page 23: ...before the pressure is stable enough for calibration Release air See Notes page 13 or 17 The effect will disappear rapidly The effect will disappear rapidly Note Always readjust the pressure after 5 1...

Page 24: ...set to the manufacturer for service please enclose a fully completed service information form Simply copy the form on the following page and fill in the required information The pressure calibration s...

Page 25: ...________________ Pressure Service request This instrument is sent for calibration please tick off ___ Calibration as left ___ Check ___ Calibration as found and as left ___ Service ___ Accredited cali...

Page 26: ...wing cleaning procedures as and when required The exterior of the instrument Clean using water and a soft cloth The cloth should be wrung out hard to avoid any water penetrating the calibrator and cau...

Page 27: ...W Range 0 6 BAR kg cm 0 10 BAR kg cm 0 16 BAR kg cm 0 40 BAR kg cm 0 100 BAR kg cm 0 300 BAR kg cm Max pressure 350 BAR 5000 PSI Min pressure 0 BAR MGC HIGH Range 0 6 BAR kg cm 0 10 BAR kg cm 0 16 BAR...

Page 28: ...p 0 to 50 C 32 to 122 F Humidity range 0 to 90 RH Protection class IP33 Wetted parts Aluminium brass stainless steel polycarbonat Leakage Max 5 over 5 minutes Reservoir volume 130 cm3 full 180 cm3 MGC...

Page 29: ...124945 01 02 10 2007 29 MGC HIGH continued Wetted parts Aluminium brass stainless steel Leakage Max 3 over 5 minutes Reservoir volume 500 cm3...

Page 30: ...2 Gauge System 4 0 10 BAR kg cm 2 MGC LOW 126319 Gauge System 4 0 300 BAR kg cm 2 MGC LOW 126320 Pump system P 142 System 1 2 3 and 4 MGC HIGH 103306T4 Gauge System 1 0 6 BAR kg cm 2 MGC HIGH 100403 G...

Page 31: ...60R120 1 0 m 3 3 ft hose with 1 4 BSP female male MGC LOW 65P175 2 0 m 6 6 ft hose with 1 4 BSP female male MGC LOW 65P180 1 5 m 4 92 ft hose with 1 4 BSP female male 60I156 5 0 m 16 4 ft hose with 1...