1 About this Supplement
This supplement is gives the information necessary to understand the
principle of operation of the Target Orbiting (TO) option fitted to your UNO
Thermometer.
This supplement must be used in conjunction with the UNO Thermometer
User Guide, supplied as standard with all UNO Thermometers.
2 UNO Target Orbiting Thermometers
The UNO Thermometers fitted with the target orbiting option are listed below:
Thermometer
Field Of View
U1 600/1600 CV - TO
100:1
U1 600/1600 CS - TO
100:1
U1 1100/2900 FV - TO
100:1
U1 1100/2900 FS - TO
100:1
U1 800/2600 CV - TO
200:1
U1 800/2600 CS - TO
200:1
U1 1500/4700 FV - TO
200:1
U1 1500/4700 FS - TO
200:1
U2 300/1100 CV - TO
100:1
U2 300/1100 CS - TO
100:1
U2 600/2000 FV - TO
100:1
U2 600/2000 FS - TO
100:1
2.1 Supplementary Specification
The following specification is in addition to that quoted in the standard UNO
Thermometer User Guide.
Half angle of conic of orbit: 2° (±10%)
Frequency of rotation:
3.0Hz (±10%)
Ambient temperature limits: 10 to 50°C / 50 to 122°F
(operation and storage)
Accuracy:
The accuracy specification for UNO
thermometers fitted with the Target Orbiting
option is given in Table 1 opposite.