Ametek LAND 4500 MkIII Installation Manual Download Page 10

Page 4

Model 4500 MkIII

Main PCB

Installation Guide

4.2  Adjusting the LED without a Modbus Connection

1)  Carefully replace the cover, ensuring that it doesn’t jam on any components
2)  When the cover is close enough, refit the keyboard connector ribbon, 

ensuring it is not offset by one pin.

 

DO NOT FASTEN THE COVER SCREWS AT THIS POINT.

3)  Apply 24 V dc power to the red 

and blue wires of Connector A.

4)  Using the keyboard, use the 

menu to display parameter 124.

5) 

VERY CAREFULLY

 remove the 

red instrument cover, taking 

care not to short-circuit any 

components inside the Model 

4500 MkIII.

6)  Adjust potentiometer PR1 until 

Register 124 has a maximum 

value not exceeding 56000.

7)  Remove 24 V dc power and 

replace the instrument cover & 

connector as before.

8)  Apply 24 V dc power to the red 

and blue wires of Connector A.

9)  Using the keyboard, use the 

menu to display parameter 122.

10) 

VERY CAREFULLY

 remove the 

red instrument cover, taking 

care not to short-circuit any 

components inside the Model 

4500 MkIII

11)  Adjust potentiometer PR2 until 

Registers 121 and 122 have a 

maximum value, with neither 

value exceeding 56000.

12)  Remove 24 V dc power.

Connector A

Connector B

Connector C

PR1

PR2

Summary of Contents for LAND 4500 MkIII

Page 1: ...Land Instruments International 2010 2017 Installation Guide Model 4500 MkIII Main PCB Part N 804320 Publication N 809352 Language English Issue 2 28 July 2017...

Page 2: ...e user documentation supplied before installing and operating the equipment The safety of any system incorporating this equipment is the responsibility of the assembler Protective Clothing Face and Ey...

Page 3: ...ent damage during transit You must include a written report of the problem together with your own name and contact information address telephone number email address etc Design and Manufacturing Stand...

Page 4: ...Model 4500 MkIII Main PCB Installation Guide...

Page 5: ...Removing the transceiver cover 1 3 Replace the circuit board 2 4 Adjust the LED intensity 3 4 1 Adjusting the LED with a Modbus Connection 3 4 2 Adjusting the LED without a Modbus Connection 4 5 Repla...

Page 6: ...Model 4500 MkIII Main PCB Installation Guide...

Page 7: ...ng System COMS CAUTION This procedure is intended for skilled instrument technicians and should be carried out in a clean workshop Take appropriate precautions to prevent damage from electrostatic dis...

Page 8: ...ke sure you can identify each of the connectors before you disconnect them If in doubt label each connector 2 Using a 3 mm Allen key remove the 6 mounting screws and remove the old circuit board 3 Fit...

Page 9: ...Connector C 2 Connect a Modbus master to the RS485 port green yellow and black wires 3 Apply 24 V dc power to the red and blue wires of Connector A 4 Observe Modbus registers 121 122 and 124 These in...

Page 10: ...instrument cover taking care not to short circuit any components inside the Model 4500 MkIII 6 Adjust potentiometer PR1 until Register 124 has a maximum value not exceeding 56000 7 Remove 24 V dc pow...

Page 11: ...ite cord seal is correctly seated before pushing the cover home 3 Secure the cover with the eight screws 4 Apply 24 V dc power and check display and keyboard functionality 6 Reconfigure and Recalibrat...

Page 12: ...erials or workmanship but due to application conditions Land Instruments International reserve the right to charge additionally for all labour on establishing this at the customer s works Warranty cov...

Page 13: ......