Caution…
Never use excessive force or tools to place or remove the
sensor as this may damage the sensor.
When the calibration is over, the STS-102 A 030 reference probe
should be removed from the insert and stored in an appropriate place,
for example in the carrying case included in the delivery.
Note…
We recommend recalibrating the STS-102 A 030 reference
probe at least once a year, more if there is any doubt about
its measuring capability.
The STS-102 A 030 reference probe may be calibrated as an
independent unit or as a system together with the calibrator.
Fig. 2
0
m
m
Reference
obe
Insert
Sensor-under-test
Reference level for
reference sensor as well
as for sensor-under-test
Reference sensor
i