Ametek IPI Mk.II Adjustment Manual Download Page 11

 

 

 

 

Conversions for Calibration in SI units. 

Since the IPI calibration points do not have to be at the exact 
PSI values stated in the procedure, when calibrating in SI units 
similar values to the following conversion can be used (500 psi 
combination range example shown). 
 
 

psi 

bar 

conversion 

bar 

calculated 

bar used 

example 

kPa 

conversion 

kPa 

calculated 

kPa 

used 

example 

500.00 

0.06894757 

34.473785 

35.00 

6.894757 

3447.3785 

3500 

437.50 

0.06894757 

30.164562 

30.00 

6.894757 

3016.4562 

3000 

375.00 

0.06894757 

25.855339 

26.00 

6.894757 

2585.5339 

2600 

312.50 

0.06894757 

21.546116 

22.00 

6.894757 

2154.6116 

2200 

250.00 

0.06894757 

17.236893 

17.00 

6.894757 

1723.6893 

1700 

187.50 

0.06894757 

12.927669 

13.00 

6.894757 

1292.7669 

1300 

125.00 

0.06894757 

8.618446 

9.00 

6.894757 

861.8446 

900 

62.50 

0.06894757 

4.309223 

4.50 

6.894757 

430.9223 

450 

0.00 

0.06894757 

0.000000 

0.00 

6.894757 

0.0000 

-12.00 

0.06894757 

-0.827371 

-0.80 

6.894757 

-82.7371 

-80 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

SPK-IPI-002, 0200007, Rev A, 9/04

 

 

 

Summary of Contents for IPI Mk.II

Page 1: ...Adjustment Manual Industrial Pressure Calibrator JOFRA IPI IPI Mk II...

Page 2: ...om the factory with all setting access unlocked or available to be changed 2 In supervisory mode each of the parameters can be locked or unlocked using the and keys Select LOC lock for those parameter...

Page 3: ...y setting a factor of 25 a 40 PSI pressure would display as 1000 40 x 25 The engineering unit displayed on the IPI will be Cust IPI Calibration Procedure Overview Calibration adjustment of the IPI is...

Page 4: ...needed to connect to the pressure standard Always make sure the hose tubing and fittings etc have a rated working pressure at or above the pressure of the unit Also it is important that there be no l...

Page 5: ...put 437 50 psi 87 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below...

Page 6: ...i 50 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressu...

Page 7: ...psi 12 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the P...

Page 8: ...PC such as Hyperterminal or PcPlus Connect the RS232 cable to stereo jack on the back of the IPI The other end of the cable should be connected to the terminal PC serial port An adapter may be needed...

Page 9: ...enu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Second point above ambient Temp Comp 7 First point below ambi...

Page 10: ...using the serial port use the VAL command Calibration Menu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Secon...

Page 11: ...94757 34 473785 35 00 6 894757 3447 3785 3500 437 50 0 06894757 30 164562 30 00 6 894757 3016 4562 3000 375 00 0 06894757 25 855339 26 00 6 894757 2585 5339 2600 312 50 0 06894757 21 546116 22 00 6 89...

Page 12: ...ment tasks from handheld field instruments to laboratory reference level bench top instruments JOFRA JF Marine Instruments A complete range of calibration equipment for temperature pressure and signal...

Reviews: