HPC50 Series

Operation Manual

 for the HPC50 Series Intrinsically Safe Handheld Pressure Calibrator

Summary of Contents for Crystal HPC51

Page 1: ...HPC50 Series Operation Manual for the HPC50 Series Intrinsically Safe Handheld Pressure Calibrator...

Page 2: ...Measuring Temperature ATMi 16 Performing a Switch Test 17 Calibrating a Pressure Transmitter 18 Enclosure 20 APMi Modules 21 ATMi Modules 22 Specifications 23 Ranges and Resolutions 28 Safety Certific...

Page 3: ...th a barometric reference BARO option all pressure readings can be displayed as gauge or absolute readings In addition to pressure measurement the HPC50 also includes inputs for mA voltage switch test...

Page 4: ...he bottom of the display Wireless Keypad All input and output connectors are placed away from the display and keypad to give maximum freedom to operate Protective Boot Required for Intrinsic Safety Mi...

Page 5: ...e different functions depending on the mode of operation Press any of the four arrow buttons to enter Navigation Mode In this mode the user can scroll through editable fields directly onscreen Navigat...

Page 6: ...disconnecting it may not be indicating true pressure Never disconnect pressure instrumentation without first relieving system pressure X To Clear the Zero Value 1 Press the zero button to activate th...

Page 7: ...ration due dates and warning period through our free CrystalControl software X Calibration Reminder Alert Enter the Cal Due date and the warning period prior Once the defined time prior is reached the...

Page 8: ...ifferent applications or tasks Up to 5 setups can be stored X General Used to view or edit the following system settings Auto Off Time HART resistor On Off Serial Number Firmware Version Part Number X...

Page 9: ...than standard is selected the displayed number will include an asterisk View Calibration Details Serial Number X To Save a Setup After you have configured the unit you can save the setup for future u...

Page 10: ...se and then press the enter button The name of the setup will appear in the upper left corner Highlight the setup then press enter DISPLAY The display is divided into two sections the upper and lower...

Page 11: ...rmation CPF o ring size and material AS568A 012 Viton 80 durometer P N 3981 For most applications CPF Fittings can be hand tightened no tools required Wrench tightening is recommended to achieve a met...

Page 12: ...condition has been reached at 80 psi i e 110 x 100 psi 30 psi 80 psi Overpressure can affect accuracy but the effect is only temporary unless the sensor has been destroyed See the Ranges and Resoluti...

Page 13: ...s 2 Press the function button for Peaks to reset the P1 P2 and APM peaks 3 Press the function button for the peaks you would like to clear The Hi Lo and Filter will all display Note If you attempt to...

Page 14: ...should have a differential reading of zero Due to the allowable error tolerance for each module the reading may not be zero The Tare function allows you to normalize both readings so that the differen...

Page 15: ...X To Measure Current 1 Select mA In For more information on working in the HPC50 display see Navigating Through a Typical Setup 2 In current mode the numerical value displayed represents the current m...

Page 16: ...ned intervals 1 Select mA Out 2 Select the appropriate mA range 0 20 4 20 3 Press the function button for Step 4 Enter the 0 and 100 mA points For example if you re sourcing to a 4 20 mA loop enter 4...

Page 17: ...the HPC50 will begin the ramp process 7 Press the function button for Stop to pause the ramp or Back to go back to the ramp setup screen MEASURING VOLTAGE The HPC50 may be used to measure voltages up...

Page 18: ...anual ATMi External Temperature Modules CAUTION ATMi modules must only be used with intrinsically safe probes provided by Ametek X To Measure Temperature 1 Connect an ATMi with RTD probe to the HPC50...

Page 19: ...he RTD and temperature switch sensor into a temperature calibrator 5 FOR A PRESSURE SWITCH Check that the vent on the pump is open and zero the calibrator if necessary Then close the pump vent The upp...

Page 20: ...sistor that can be selected if desired With the settings button choose General to select HART resistor On or Off X Percent Error Function The HPC50 can be programmed to calculate a deviation or error...

Page 21: ...ay will indicate the error as well as the actual mA value The lower display will show the measured pressure 4 20 Flow 0 20 Flow 0 4 0 25 5 1 25 50 8 5 75 13 11 25 100 20 20 X Scaling Function The HPC5...

Page 22: ...67 Keypad and Labels UV Resistant Silicone ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS 1 4 INCH MNPT ADAPTER 1 4 INCH MBSP ADAPTER M20X1 5 ADAPTER 29 1 1 15 26 1 1 03 32 6 1 30 176 3 6 94...

Page 23: ...m specification applies only where specified Not recommended for continuous use at high vacuum Refer to XP2i DP data sheet for gauges that are intended for continuous high vacuum use The BARO option f...

Page 24: ...HPC50 SERIES CABLE 6 PIN CONNECTION FOR REMOTE TEMPERATURE PROBE 63 3 2 49 27 1 06 Class A Class B Temperature C HPC50 ATMi Uncertainty Class A Uncertainty HPC50 ATMi Class A Uncertainty Class B Unce...

Page 25: ...eturn to a zero reading or return to the same ambient barometric reading X Gauge Pressure psi 18 to 28 C 0 to 30 of Range 0 01 of Full Scale 30 to 110 of Range 0 035 of Reading Vacuum 0 05 of Full Sca...

Page 26: ...function 0 05 of static line pressure reading Barometric Reference BARO Accuracy 0 00725 psi 0 5 mbar Range 10 153 to 15 954 psiA 700 0 to 1100 0mbarA Units and Resolution psi 0 001 inHg 0 001 mmHg 0...

Page 27: ...ls Electrical All accuracy specifications include all effects of linearity hysteresis repeatability temperature and stability within the specified operating temperature range for one year Note Exposur...

Page 28: ...Accuracy 0 015 of rdg 2 mV Range 0 to 28 VDC Maximum Allowable Voltage 30 VDC Resolution 0 001 VDC X Switch Test Switch Type Dry Contact Closed State Resistance 1K Open State Resistance 100K Sample R...

Page 29: ...ll MN1500 T3 20 to 40 C T3 20 to 50 C Energizer EN91 Replace batteries with approved type in non hazardous locations only Energizer is manufactured by Energizer Holdings Inc and the Eveready Battery C...

Page 30: ...ressure psi kg cm2 inHg inH20 mmHg mmH20 kPa bar mbar MPa 15PSI 3 0 x 0 0001 0 00001 0 001 0 01 0 01 0 1 0 001 0 00001 0 01 1BAR 3 0 x 0 0001 0 00001 0 001 0 01 0 01 0 1 0 001 0 00001 0 01 100KPA 3 0...

Page 31: ...cation where external conditions are conducive to the buildup of electrostatic charge CERTIFICATIONS The HPC50 Series has been tested and certified to comply with a variety of international standards...

Page 32: ...alled in a location where external conditions are conducive to electrostatic charge build up Fran ais French Si l quipement est utilis d une mani re non sp cifi e par le fabricant la protection fourni...

Page 33: ...gas and fluid is acting like a spring At higher pressures above 2000 psi typically this may eventually diminish as the gas dissolves into the fluid X Solution Evacuate all tubing with a vacuum pump b...

Page 34: ...full scale This level of accuracy requires the use of piston deadweight gauges or very high performance pressure controllers There are no internal potentiometers Each of the HPC50 s sensors has a span...

Page 35: ...rable and high speed Micro USB cable enables data transfer in the most convenient form Included as standard P N 3009 Hard Carrying Case Aluminum carrying case with molded eggshell foam interior P N 28...

Page 36: ...material and workmanship under normal use and service for one 1 year from date of purchase to the original purchaser It does not apply to batteries or when the product has been misused altered or dam...

Page 37: ...2020 Crystal Engineering Corporation 708 Fiero Lane Suite 9 San Luis Obispo California 93401 8701 6486 D 2009...