background image

5-4   |   3050 AM Moisture Analyzer      

Echo (A)

Echo implements a loop back function to test the integrity of the serial 
communication. Any text transmitted by the master is returned.

Example

:

Master:

  

>99ABCDEFG4E[CR]

Slave:

   

  AABCDEFG1D[CR]

Bad Command (B)

This command is not recognized by the slave. It will therefore result in a 
type 01 failure response.

Master:

  

>99BB4[CR]

Slave: 

 

  N01[CR]

Acknowledge (C )

Requests a simple acknowledge from the slave. Useful to determine if 
there is a device at that node address.

Master:

  

>99CB5[CR]

Slave:

   

  A[CR]

Read Data (F)

The data field of this frame contains the handle for a system variable. See 
Figures 5-7a and 5-7b for a list of defined handles for system variables. 

Example

:

Master:

  

>34F0512[CR]

Slave:

  

A1234540[CR]

Summary of Contents for 3050-AM

Page 1: ...3050 AM Moisture Analyzer User Manual Process Instruments 455 Corporate Boulevard Newark DE 19702 PN 305804901 Rev YA Configurator Version 3 0...

Page 2: ...ING USA Delaware 455 Corporate Blvd Newark DE 19702 Tel 302 456 4400 Fax 302 456 4444 USA Oklahoma 2001 N Indianwood Ave Broken Arrow OK 74012 Tel 918 250 7200 Fax 918 459 0165 USA Pennsylvania 150 Fr...

Page 3: ...1 3 Gas Flow 1 3 Sample system 1 3 Sample flow 1 4 Zero gas module 1 5 Zeroing the Analyzer 1 5 Dew Frost Point Measurements 1 6 Chapter 2 Specifications Chapter 3 Installation and Start Up Analyzer...

Page 4: ...Outputs 4 3 Adjust Offset 4 3 Zero 4 3 Abort 4 3 Alarm Key 4 4 Range Key 4 5 Config Key 4 6 Display 4 6 Communication 4 7 Sample Gas 4 8 Custom gas 4 8 Clock 4 9 Zero Schedule 4 10 System Test 4 12 C...

Page 5: ...s required 7 10 Procedure for removing the electronics module 7 10 Proportional Solenoid Valve PSV PCB 7 12 Analog Interface PCB 7 12 Removing the analog interface PCB 7 12 Replacing the analog interf...

Page 6: ...tions and ground checks Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection originally provided by the equipment Grounding Instrument grounding is m...

Page 7: ...t and to conserve our natural resources AMETEK recommends that you arrange to recycle this prod uct when it reaches its end of life Waste Electrical and Electronic Equipment WEEE should never be dispo...

Page 8: ...c capacitor is a technique allowing high frequency shield bonding while avoiding the AC ground metal connection In the case of shielded cables the drain wire or braid connection must be kept short A t...

Page 9: ...Off or the Area is Known to be Non Hazardous Risque d explosion Avant de d connecter l quipement coupez le courant o vous assurez que l emplacement est design non dangereux All input and output wiring...

Page 10: ...uct Product Name Model 3050 AM Moisture Analyzer Series Conforms to the following standards Electromagnetic Compatibility Directive 89 336 EEC EN 61326 1 Radio Frequency Emissions EN 55022 CISPR 22 Co...

Page 11: ...umentation accompanying the equipment being returned declaring that the goods being returned for repair are American goods the name of the firm who purchased the goods and the shipment date The warran...

Page 12: ...xii 3050 AM Moisture Analyzer This page intentionally left blank...

Page 13: ...or The calibrated range of the analyzer is 0 to 100 PPMV with a trend indication for measured concentrations to 1000 PPMV The QCM sensor technology used in this analyzer is compatible with inert gases...

Page 14: ...xternal PC 4 line x 20 character LCD display that provides moisture data system operating parameters system messages and alarm conditions An 18 button keypad on the front of the unit that is used to s...

Page 15: ...gas flows through the sensor and a second measurement of the sensor frequency is recorded The difference between these two frequency measurements is known as the delta frequency The value is a functio...

Page 16: ...ample path and the reference path The sample path flows through the first propor tional valve PSV1 and the reference path flows through a dryer and PSV2 The two proportional valves PSV1 and PSV2 contr...

Page 17: ...ough the dryer providing a dry reference 10 PPBV for the analyzer By providing this dry reference gas the zero gas module enables the offsets present in the system e g electronic offsets offsets cause...

Page 18: ...y the Model 3050 AM is restricted for many sample gases Many gases liquefy at relatively low pressures so calculating a dew frost point temperature for these compounds has no value Therefore the 3050...

Page 19: ...sensor assembly Contact the factory to confirm compatibility with other gases Read the Dryer Warning in the Special Warnings and Information section in the beginning of this manual for more informatio...

Page 20: ...0 mA DC analog output RS 232 RS 485 serial port Alarms Alerts Three independent contact closures of 30 VAC or 60 VDC maximum 10 VA or 1 A maximum resistive for system alarm moisture concentration alar...

Page 21: ...ounting 19 inch rack Dimensions Width 48 cm 19 inches Height 13 cm 5 25 inches Depth 42 cm 16 5 inches Net Weight 9 8 Kg 22 lb Approvals and Certifications MET Certified to UL1604 and CSA213 General S...

Page 22: ...2 4 3050 AM Moisture Analyzer This page intentionally left blank...

Page 23: ...Class I Division 2 Groups A thru D or Non hazardous locations only This chapter contains information on the installation and setup of the 3050 AM Moisture Analyzer including Analyzer Requirements Spa...

Page 24: ...entilation Figures 3 1a b and c It should be located as close to the sample tap as possible Figures 3 1b Model 3050 AM dimensions rear bottom Sample In 20 50 PSIG Exhaust Out Atmospheric Pressure Inst...

Page 25: ...Installation Start Up 3 3 Figures 3 1c Model3050AMdimensionstopand sides...

Page 26: ...ir 70 to 100 PSI that is free of oils and other contaminants for valve operation N2 can be used as an alternate If there is any doubt about the cleanliness of the air supply con nect it through oil ai...

Page 27: ...elded pair External Sample Tubing Sample input tubing length should be as short as possible For optimum performance sample tubing should be thermally con trolled to maintain a constant sample input te...

Page 28: ...to operate internal diaphragm valves 1 4 inch Swagelok fitting 4 RS 232 Serial Port User serial communications interface Serial wiring supplied by user Observe caution regarding use in hazardous loca...

Page 29: ...RS 485 termination switch is on if the analyzer is the only analyzer on line or the last in a chain 12 RS 485 Serial Port User serial communications interface Serial wiring supplied by user 13 Analog...

Page 30: ...avoid damage to the device 1 On the right side of the rear of the analyzer is the dryer acess com partment Figure 3 4 2 Loosen the four captured screws on the door of the compartment to remove the do...

Page 31: ...finger tight 9 Tighten the nut an additional 1 8 turn 10 Repeat steps 6 through 9 for the second dryer which you will mount in the bottom or zero dryer position Figure 3 4b 11 Replace the dryer compa...

Page 32: ...d period of time reinstall the caps using new gaskets 2 Connect the sample input tube to the analyzer sample inlet 1 4 inch VCR fitting 6 Be sure to use a new gasket 3 Connect the exhaust tube to the...

Page 33: ...rocedure to prevent damage to malfunction of the analyzer 1 Ensure that the area is non hazardous and the analyzer has been de energized and power disconnected 2 Remove the terminal cover on back pane...

Page 34: ...ief cap with compression bushing 1 Figure 3 2 and pull out the power cord 5 Prepare the new power cord using the original cord as a guide remove outer jacket measure length of wires strip lengths of w...

Page 35: ...t 13 see Figures 3 10a and 3 10b and alarm and data valid terminals 14 on the interface board to the analyzer through the metallic signal conduit 2 Figure 3 2 ON ON TB3 TB2 TB1 ANALOG OUTPUT DATA VALI...

Page 36: ...onnection analog output and alarm data valid contacts share the same cable RS 485 communications Turn on the RS 485 termination switch 11 Figure 3 9 if communi cating with one analyzer or the last in...

Page 37: ...ions 4 to 20 mA 1 2 3 4 5 6 7 8 9 10 Meter 100 to 500 ohms TB2 DC Supply 24 Volts Figure 3 10b Externally powered 4 20 mA output connections 1 2 3 4 5 6 7 8 9 10 Meter 100 to 500 ohms TB2 Figure 3 10a...

Page 38: ...on demand zero settings and initiate a zero cycle ALARM Used to define alarm output settings ANALOG RANGE Used to scale the 4 to 20 mA output in pro portion to the moisture concentration TEST CONFIG...

Page 39: ...ancel To exit without saving when entering data Any changes made will not be saved To go back through previous levels one level at a time until the display returns to the normal opera tion screen when...

Page 40: ...uld allow 5 to 7 days for the analyzer to dry down and stabilize During this period the concentration value shown on the display and the data recorded from output should show a steadily decreasing tre...

Page 41: ...ry day at 1 AM The analyzer will spend 50 of the time on the sample gas and 50 of the time on the internal zero These settings enable the analyzer to more effectively dry all of the surfaces within th...

Page 42: ...3 20 3050 AM Moisture Analyzer This page intentionally left blank...

Page 43: ...nc tion keys numbers minus sign and decimal point you can set the param eters for your analyzer 5 7 ppmv Sampling High Concentration 5 0 ppmv Limit Line 1 Concentration Line 2 Status message Line 3 Al...

Page 44: ...n 116 Helium Hydrogen Krypton Neon Nitrogen Oxygen SF6 Xenon Baud Rate RS 485 Mode Two Wire Four Wire Node Address Communication Font Enter Value 0 or 1 Enter Address Sunday Monday Tuesday Wednesday T...

Page 45: ...to change during a zero Adjust Offset Enable Select this function to adjust the offset at the end of a zero cycle Disable Select this function if you do not want to adjust the offset Zero This functio...

Page 46: ...high limit value in moisture units selected Low Limit Enter the low limit value in moisture units selected Enable Alarm Select Enable or Disable to activate the moisture concentration alarm relay Fig...

Page 47: ...nal to the moisture readings Maximum Conc Enter the moisture reading equivalent to 20 mA Minimum Conc Enter the moisture reading equivalent to 4 mA Example To scale the analog output to represent a 0...

Page 48: ...to 9 The default is 5 Back Light Select On or Off The default is ON Font This is a factory set parameter that identifies the type used for the display on the ana lyzer Config Display Contrast Back Lig...

Page 49: ...600 19200 Two Wire Four Wire Node Address Enter the address for RS485 serial communications 0 240 Address 240 is reserved see the Special Addresses section in Chapter 5 for details Baud Rate Select th...

Page 50: ...3050 AM to adjust for the molar heat capacities of different gases Nitrogen is used as the reference gas for the K factors and its value is defined as 1 000 Please consult AMETEK Process Instruments...

Page 51: ...Value Enter Value Enter Value Sunday Monday Tuesday Wednesday Thursday Friday Saturday Year Enter the 4 digit value for the year Day Enter the 2 digit value for the day Day of the Week Select the day...

Page 52: ...ly Monthly Day of the Week Select the day of the week from the list Day of the Month Enter the 2 digit value for the day of the month Hour Enter the 2 digit value for the hour of the day using a 24 ho...

Page 53: ...in the reference leg of the analyzer The remaining dryer capacity is reset when a new number is entered Moisture Units Select Units NOTE NOTE The units for the moisture readings can be selected from a...

Page 54: ...Alarms Tests the alarm contacts by toggling them open close Test Analog Output Tests the analog output by allowing the user to set outputs to 4 12 or 20 mA Test Serial Communications Performs a loop...

Page 55: ...ag is a carriage return 0D The Node address is a two byte hexadecimal ASCII value in the range 00 through EF Addresses F0 through FF are reserved The command byte is a single ASCII character The Data...

Page 56: ...a field is seen in Figure 5 2 The message format for slave acknowledge without a data field is seen in Figure 5 3 The message format for slave failure response is seen in Figure 5 4 Figure 5 2 Slave r...

Page 57: ...ges with node address F0 You must be careful to prevent multiple slaves from receiving commands with the broadcast address as this will cause a bus contention Defined Commands master to slave There ar...

Page 58: ...cognized by the slave It will therefore result in a type 01 failure response Master 99BB4 CR Slave N01 CR Acknowledge C Requests a simple acknowledge from the slave Useful to determine if there is a d...

Page 59: ...xample Master 34H05123 40C CR Slave A CR Write Cell EEPROM L Copy cell serial number and cell calibration coefficients into cell EEPROM Example Master 34LB3 CR Slave A CR Quit Q Zero cycle is terminat...

Page 60: ...e returned a simple acknowledge consisting of an ASCII A will be sent Example Slave A CR Acknowledge with Data If a command issued by the master requires data in response an Acknowl edge with Data res...

Page 61: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes Failure Code Description 0 Unknown Command Byte 02 Check Sum 05 Parameter Out of Range...

Page 62: ...ded software version String 1 00 HoldOut 69 45 Hold out during zero and verification Integer 1 hold ZeroOnly 72 48 Start zero only flag Integer 1 zero only EnableAlarm1 73 49 Enable or disable concent...

Page 63: ...t time read Integer 2003 Date 106 6A Analyzer current date String 02 24 03 Time 107 6B Analyzer current time String 08 45 SystemState 108 6C Alarm status word Integer 1024 DataState 110 6E Data ready...

Page 64: ...5 10 3050 AM Moisture Analyzer This page intentionally left blank...

Page 65: ...let and outlet pressures Call AM ETEK service if problem persists Battery Alarm Low battery voltage detected Yes Replace battery or call AM ETEK service Reference Gas Alarm Delta frequency is less tha...

Page 66: ...6 2 3050 AM Moisture Analyzer This page intentionally left blank...

Page 67: ...ted from the analyzer before attempting to remove or replace any of the parts CAUTION Sample System Sample System Connections Bracket Zero Module Dryer Po we r Su pp ly 24 VD C PCB Top Fan PSV Top REA...

Page 68: ...d Sample System Connections Bracket Solenoid Solenoid Capillary Zero Module Exhaust Sample In Fan Pressure Transducer Sensor Metal Cans 1 16 Fittings Valve Manifold Pressure Transducer Figure 7 2 Mode...

Page 69: ...being careful to free the ribbon cable attached to it from the other wires that pass through to the electronics compartment 5 Disconnect the ribbon cable from the analog interface board and dis card t...

Page 70: ...Figure 7 4 3 Disconnect the capillary and remove it by loosening both nuts and gently pulling the tubing out of the stream switching manifold block Installing the capillary 1 To install the capillary...

Page 71: ...be intact at this time you will need to rest the sample manifold plate on top of the analyzer to perform the valve replacement 5 Disconnect the 1 16 inch fittings on the inlet and outlet located on e...

Page 72: ...then reconnect the 1 16 inch fitting at the inlet and outlet of the valve gently pushing the tubing in and securing with the nut 3 Place the sample manifold plate back in the oven and secure with the...

Page 73: ...top position and the zero dryer is located in the bottom position 5 Taking one dryer at a time use a 7 64 hex wrench to remove the mounting screws that pass through the top of the dryer body and set...

Page 74: ...ey are finger tight 3 Install the dryer mounting screws to secure the dryer to the bracket 4 Tighten both of the VCR fittings so that the nuts are 1 8th of a turn past finger tight 5 Place the dryer c...

Page 75: ...equence is complete Press TEST CONFIG Scroll down to Production Codes and press Enter Select Reference Dryer and press Enter to enter the production code Press Enter Press Cancel several times to reve...

Page 76: ...s module Use Figure 7 7 as a reference for locating the items described in this proce dure Locate the four 4 Phillips head screws on the electronics module Using the appropriate tool remove the mounti...

Page 77: ...s Power Supply 24 VDC P C B T o p P il o t V a lv e M a n if o l d PSV PilotValve M anifold PSV Power Supply 24 VDC Power Supply Pilot Valve Manifold P C B T o p M C U T o p Pilot Valve Receptacle Plu...

Page 78: ...PSV board in place Detach as described above Unplug the connectors attached to the Analog PCB board Figure 7 7 and Figure 7 9 Carefully lift the board until it slides free of the connector Replacing t...

Page 79: ...Power Figure 7 10 MCU PCB connections P2 RS 485 P3 1 Reference Valve P3 3 Sample Valve P4 RS 232 Figure 7 8 Analog interface board connections P6 3 P6 4 C2 U1 P Q D1 D2 Q2 305517 Rev AMETEK C1 AM AP J...

Page 80: ...old plate Figure 7 7 and Figure 7 11 2 Disconnect the pilot valve receptacle plugs Figure 7 11 3 Use a 1 16 slotted screwdriver to remove the two 2 screws at the side of the manifold Figure 7 11 The m...

Page 81: ...Reinstalling the Electronic Module Once you have serviced the various components in the electronics mod ule reassemble the analyzer 1 Carefully reposition the electronic module above and to the front...

Page 82: ...6 3050 AM Moisture Analyzer P1 Heater Wire P3 120 Volt P4 RS 485 P5 System Alarm Board Connections User Interface Board 120 Volt System Alarm RS 485 Heater Wire 120 Volt System Alarm RS 485 Circuit Si...

Page 83: ...517901 Pilot Valve Manifold Assembly 583011901S Sample and Reference Proportional Valves 305705901 Power Supply 15 V 5V T460 048 Power Supply 24 V 271768001 Dryer 583072901S Oven Fan Large 583090901 O...

Page 84: ...7 18 3050 AM Moisture Analyzer This page intentionally left blank...

Page 85: ...the configurator software CD into CD ROM drive 2 The installation program should begin to run immediately If it does not start automatically click RUN from the Windows Start menu Type the drive lette...

Page 86: ...setup any changes to the analyzer communications pa rameters must be made using the Device Communications tab These changes must be made before you make any changes to the computer serial port setting...

Page 87: ...zer Description Type in a description for the analyzer Save Configuration button Save the analyzer s internal parameters to a file The Save As dialog box opens so that you can name and save the file R...

Page 88: ...or serial number and the dryer code are also displayed In the lower right hand corner above the Help button is the Configurator software version number Saving Your Settings To save settings on the Gen...

Page 89: ...erial communication cable that is being used Address Type the network address for the analyzer being con nected Two Wire RS 485 Click if you are using a 2 wire cable Four Wire RS 485 Click if you are...

Page 90: ...ement Process Pressure Enter the process pressure in kPa 4 20 mA Output Set up your analog output range 20 mA Enter your high analog output limit 4 mA Enter your low analog output limit Hold during Ze...

Page 91: ...cle Not checked The analyzer verifies only Set Clock Time Synchronization Click Set Clock to synchronize the clock within the analyzer with the PC The Time Synchronization box opens with the PC time a...

Page 92: ...hour clock Example 1 00 PM 13 hours Weekly Select the day of the week from the drop down list under Day of the week at the bottom of the box You must also enter a time of day 1 through 24 in the Hour...

Page 93: ...en opens so that you can browse to where you want to save the file The file format is Excel compatible CSV format Concentration readings data for the last 24 hours can be written to this file from the...

Page 94: ...s Save button The file format is csv which is Excel compatible All data displayed on the Monitor page will be stored in the specified file Figure 3 10b Monitor tab Save As Off Data collection will ter...

Page 95: ...main shutoff valve at the source 2 Turn on the AC power at the source 3 Once the sample flow starts the analyzer will be in normal operating mode Allow analyzer to warm up for 45 minutes Dry Down All...

Page 96: ......

Reviews: