1.2 Specifications - LSP
HD
Series
1.2.1 Common specifications
Ambient temperature range
5 to 60 °C / 41 to 140 °F (in specification)
5 to 70 °C / 41 to 158 °F (operating)
Alignment
Class 2 Laser indicating scan plane and extent
Laser scan rate 1 Hz
Internal Ambient Temperature Sensor
(IATS)
-10 to 90 °C / 14 to 194 °F
Emissivity range
(remote or internally set)
0.20 to 1.00
Sealing
IP65
EMC
LVD
EN 61326 Class A [emissions & immunity]
EN61010-1
Vibration (Any axis)
2g [10 to 30 Hz]
3g [30 to 300 Hz]
1.2.2 LSP
HD
1 series specifications
LSP
HD
10
LSP
HD
11
Operating wavelength
1µm
1µm
Temperature range
600 to 1400 °C
1112 to 2552 °F
700 to 1500 °C
1292 to 2732 °F
System Accuracy
(1)
(5 to 95% of range)
± 2 °C / 3.6 °F
Focus distance
(2)
1m / 39.7in to infinity (continuously
adjustable)
Field of View
(static to 95% of radiance)
500:1 (0.002 radians)
Repeatability
< 0.5 °C / 0.9 °F
Temperature Resolution
(3)
< 5% of range
(noise)
< 15% of range
< 30% of range
Above 50% of range
≤ 6 °C / 10.8 °F
≤ 6 °C / 10.8 °F
≤ 4 °C / 7.2 °F
≤ 1 °C / 1.8 °F
Scan Angle
(4)
80°
Scan Rate
Factory set
User adjustable
50 Hz
10 to 150 Hz in 1 Hz steps
Speed of response
(Tau - linear output on rising edge)
≤ 1 µs
Ambient Stability
deltaT[indicated] / deltaT[ambient]
≤ 1°/10°
(1) Includes error correction
(2) Focus is factory set to 2m
(3) 4 line averaging below 15% of range
(4) Scan angle may be set in the range 40 to 80° by the PC software