Page 23
User Guide
Thermometer
Notes on DPU interconnections (Fig. 18)
‘
To indicator
’
The indicator must have a ‘floating’ input i.e. a differential or isolated
input.
‘
Reset control
’
The reset control must be electrically isolated from the indicator.
When used with an UNO, only the connections shown in Fig. 18 must
be made to the DPU. Do not make any other connections to the DPU.
DPU Pin Nº Cable Colour
Function
UNO to DPU
connections
1
Yellow
4 to 20mA linear temperature signal drive
2
Blue
4 to 20mA linear temperature signal return
3
White
CMD (Command) input
4
Screen
Screen
5
Red
23 to 48V, <200mA, dc power (positive)
6
Black
23 to 48V, <200mA, dc power (negative)
7
Green
CMD (Command) input
8
N/C
No connection
External connections
to DPU
9
Brown/Red
AC Mains Live
10
Black/Blue
AC Mains Neutral
11
White
CMD (Command) input*
12
Green
CMD (Command) input*
13
Screen
Screen
14
Yellow
4 to 20mA linear temperature signal drive
15
Blue
4 to 20mA linear temperature signal return
16
N/C
No connection
Notes
Recommended output load resistance: 100
W
Maximum output load resistance:
400
W
*CMD (Command) input:
Voltage-free switch closure
Switch open = PEAK PICK
Switch closed = TRACK (or RESET)
Fig. 19 UNO to DPU interconnection schedule