6.4 Control Thermocouple
The thermocouple which controls the furnace temperature is situated close to
the furnace heating elements, which will be at a higher temperature than the
target cavity. The difference between controller set point temperature and
target cavity temperature can be as high as 20°C / 40°F at 1500°C / 2700°F.
In these cases, if a target temperature of 1500°C / 2700°F is required, it is
necessary to adjust the controller setpoint to 1520°C / 2740°F.
The temperature indication on the controller must not be used as an accurate
measurement of target cavity temperature.