GENERAL SAFETY INFORMATION
1
6
DR7000 SERIES
www.drexelbrook.com
EDO 11-10-112, DR7000-LM ATEX, Issue #2
1.5 ATEX nameplates
Figure 1-1: Ex ia / Ex iaD nameplate (4...20 mA output)
Figure 1-2: Ex ia / Ex iaD nameplate (PROFIBUS PA or FOUNDATION
™
Fieldbus output)
1
ATEX certification agency code
2
Equipment category (explosive atmosphere - gas)
3
Types of device protection including approved Gas Groups (IIA, IIB or IIC) and temperature classes (T6...T3)
4
Equipment category (explosive atmosphere - dust)
5
Types of device protection (explosive atmosphere - dust)
6
Degree of ingress protection (if fitted with the appropriate cable glands) and maximum surface temperature
7
Intrinsically-safe circuit data
8
Cable entry type and size (Aluminium housing: M26×1.5, M20×1.5,
½
NPT or G
½
; Stainless steel housing: M25×1.5,
M20×1.5,
½
NPT or G
½
)
!