1
Reboot system.
2
If error remains, call service.
1
Call service for recalibration.
1
Select other instruction, or change mode (End, Run or
Pause).
1
Reboot system.
2
If error remains, notify instrument support.
1
Low pressure limit can not be set higher than high pressure
limit. Reset limit.
1
Low pressure limit for warning can not be set higher than
high pressure limit for warning. Overlapping ranges due to
hysterersis are not allowed. Reset limit.
1
Release pressure in system before setting analogue output
voltage.
1
Call instrument support.
1
Full scale pressure must be greater than zero. Reset value.
3636 Value must be
greater than 0
3620 Error hardware
notify instr suppor
3610 Discon pressure
then set analog V.
3605 Low lim>hi-lim
Hysteresis
3600 Low lim>hi-lim
3624 Error softwar
notify instr support
3622 Error softwar
notify instr support
3621 Error softwar
notify instr support
3510 Error softwar
notify instr support
3411 Error softwar
notify instr support
3410 Error softwar
notify instr support
3401 Error softwar
notify instr support
3400 Error software
notify instr support
3300 Error software
notify instr support
88 Not allowed in
current mode.
86 ERROR in EEPROM
Recalib. all values
85
Failed to set offset
Pump P-920
User Manual 18-1125-54 Edition AB
26
5
Trouble-shooting