Monitor pH/C-900
User Manual 18-1120-06 Edition AC
23
5 Select main menu
Setup
, press OK.
6 Select sub menu
Setup Conductivity
,
press OK.
7 Select sub menu
Setup Adjust Cond
,
press OK.
8 A warning message
will be displayed,
press OK.
9 Enter the
theoretical
conductivity value
according to the
graph, press OK.
The new cell
constant is
automatically
calculated and
updated.
B.2.3 Entering the conductivity cell constant
After replacing the flow cell, the cell constant has to be set. (The cell
constant is shown on the packaging).
1 Select main menu
Setup
, press OK.
2 Select sub menu
Setup Conductivity
, press OK.
3 Select sub menu
Setup Adjust Cell Constant
, press OK.
4 A warning message will be displayed, press OK.
5 Enter the cell constant, press OK. The new cell constant is
updated.
When entering the cell constant from UNICORN select
System
Control:System:Calibrate
and select
Cond_Cell
.
Setup Adj Cell Const
(34.4/cm) 35.5
Warning! This will
change cell calib
Setup Adj Cell Const
(34.4/cm)
Setup Conductivity
Setup Adjust Cond
(83.53mS/cm)
Setup Adjust Cond
(83.53mS/cm) 86.60
Warning! This will
change cell calib
Setup Conductivity
20
25
30
77
80
85
90
95
97
Temperature (°C)
Conductivity (mS/cm)
Conductivity of 1.00 M NaCl at 20–30°C
Reference information
B
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 23 (1,1)
Summary of Contents for AKTA design pH/C-900
Page 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Page 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Page 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Page 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Page 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...