1.2     SYSTEM FLOW DIAGRAM 

                                                  

  

                                                                  Figure 1.1 

Summary of Contents for RAPID RESPONSE

Page 1: ...98 2024 Rev A RAPID RESPONSE WATER SYSTEM FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Road Dayton OH 45414...

Page 2: ...98 2024 Rev A...

Page 3: ...YSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 7 2 4 ENDOTOXIN FILTER 8 3 0 SYSTEM START UP 9 3 1 OPERATION 9 3 2 BYPASS PROCEDURE 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANC...

Page 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Page 5: ...ns Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours emergencies...

Page 6: ...ures and voltage draws please refer to listed component manuals as listed below For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wal...

Page 7: ...4 98 2024 Rev A 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Page 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Page 9: ...ght of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to the wall 3 Connect tubing from the SRO product flow meter o...

Page 10: ...ansformer into the receptacle and open the inlet and sample port valves 2 3 STORAGE TANK 1 Locate the storage tank on a firm level foundation and anchored securely 2 Connect the Wall Mount DI product...

Page 11: ...torage tank to the SRO controller and lock in place Figure 2 4 2 4 ENDOTOXIN FILTER 1 Locate and install the Endotoxin Retentive Filter assembly on or near the wall 2 Connect the 3 4 pump discharge ho...

Page 12: ...8 0160 For 00HC 0090 PRV Healthcare Storage Tank refer to Manual 098 0001 For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 3 1 OPERATION Once the system has been started the SRO will continue...

Page 13: ...ce your system components Important note If you have a return flow line coming back to the storage tank when using the bypass the returning water from your process will overflow the tank as the distri...

Page 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Page 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Page 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Page 17: ...ees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes from...

Page 18: ...15 98 2024 Rev A Figure 5 1...

Page 19: ...END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endotoxin...