Start-Up
Installing the resin pack
1. With the isolation valves turned off, loosen the tube fitting on the bottom of the
deionizer housing. Remove the tubing from the fitting.
2. Using the filter wrench, loosen and remove the
lower deionizer body by turning it counterclockwise.
Verify that the O-ring stays in the O-ring groove.
3. Install the resin pack into deionizer body.
4. Attach the deionizer body back onto the cap and reconnect the tube fitting.
5. Remove the tubing from the deionizer outlet isolation valve and place it in a
bucket or direct to a drain.
6. Open the inlet isolation valve slowly to start water flow to the deionizer.
7. Allow water to run through the deionizer until all air is purged from the system.
After all of the air is removed reinstall the outlet tubing to the outlet isolation
valve.
8. The quality monitor may glow red for a couple of minutes before changing to
green.
Open the drain valve on the bottom of the housing.
Remove the resin pack from the box and plastic bag.
Stretch the resin pack over your arm to elongate the
pack, and feed it into the housing.