background image

ATP-100 Instruction Manual 

Tuning Pulser 

PULSE RATE and DUTY CYCLE 

The ATP-100 has two internal controls that adjust Pulse rate and duty cycle. As 
indicated by the names, these controls adjust the rate at which the transmit 
keying pulses occur (pulse rate), and the ratio of on- to-off time (duty cycle).  
The left control adjusts the Pulse rate, while the right control adjusts the duty 
cycle. 
 
Some care must be used to insure PULSE RATE is slow enough, and the DUTY 
CYCLE is long enough, to be within the keying response limits of the exciter 
and meter. If the pulse rate is too fast and/or the duty cycle is too long,  the 
pulses will blur into one long steady signal.  
 
The typical exciter has a leading edge rise time and tailing edge decay time of a 
few milliseconds. This stretches the pulse duty cycle out, making the actual RF 
envelope have a longer duty cycle than the actual keying waveform from the  
 
ATP-100. Because of this effect, the Ameritron ATP-100 Tuning Pulser  may 
actually produce a 100% duty cycle waveform at high (clockwise) PULSE 
RATES with longer (clockwise) on-time DUTY CYCLE settings.   
 
If the pulse rate is too slow and/or the duty cycle too slow, peak power will not 
be reached. All exciters have a delayed response to the leading edge of the 
keying waveform. This delay may prevent full peak power from being reached if 
the duty cycle is too short. Most peak meters have limited response time, this 
causes lower power readings with short duty cycle pulses.  
 
Another problem is nearly all meters have limited storage time, causing the 
meter to "fall-back" during the time interval between RF pulses. To obtain the 
most accurate 

meter readings

, both PULSE RATE and DUTY CYCLE should 

be set far enough clockwise to allow maximum peak power to be indicated on a 
scope or peak reading meter.  
 
Some peak meters do not read true peak power at all, instead reading something 
less than the true peak power. If your meter reads less power when the pulser is 
used, compared to the power when a steady or near steady carrier is produced, 
your meter is probably not a good peak reading meter. Nearly every radio and 
amplifier produces more peak power than steady carrier power, because of ALC 
response time and power supply voltage sag under load. Too much duty cycle 
on-time can cause unwanted component heating. 
 

Summary of Contents for ATP-100

Page 1: ...cuit The Ameritron ATP 100 Tuning Pulser can also be used as a valuable troubleshooting and diagnostic aid The Ameritron ATP 100 Tuning Pulser can be used in performance tests of QSK systems wattmeter...

Page 2: ...ech processing increases the average power pushing the short term average power of normal speech to 30 or more During normal voice operation most equipment heat is generated by the quiescent current r...

Page 3: ...longer clockwise on time DUTY CYCLE settings If the pulse rate is too slow and or the duty cycle too slow peak power will not be reached All exciters have a delayed response to the leading edge of the...

Page 4: ...ck hazard and potential of accidentally keying the rig if the cabinet of the Ameritron ATP 100 Tuning Pulser or the operator comes in contact with grounded equipment WARNING Never use the ATP 100 with...

Page 5: ...ol to the rated drive power of the amplifier Place the amplifier into operate mode and depress the pulse tune button on the Ameritron ATP 100 Tuning Pulser Tune the amplifier for peak output power Not...

Page 6: ...ATP 100 Instruction Manual Tuning Pulser 6 Schematic...

Page 7: ...100 R1 Resistor 47K OHM 100 4470 R6 R7 Resistor Trimpot 500K OHM 133 5500 C1 C3 Capacitor 01uF 200 0416 C2 C4 Capacitor 1uF 200 0754 D1 D2 D3 Diode 1N4148 300 0140 D4 Diode 1N4007 300 0266 Q1 FET 2N70...

Page 8: ...measuring the single tone power with a barefoot exciter running off of a regulated supply and adjusting an oscilloscope to read near full screen with that carrier By activating the ATP 100 with the PU...

Reviews: