background image

ATP-100 Instruction Manual 

Tuning Pulser 

Ameritron ATP-100 Tuning Pulser  

The Ameritron ATP-100 Tuning Pulser relieves temperature related stress on 
amplifiers, tuners, and dummy loads while allowing proper system adjustments. 
It allows amplifiers to be properly adjusted without the use of long steady 
carriers. 
 
The Ameritron ATP-100 Tuning Pulser allows tuning of an amplifier with full 
required drive from the transceiver without over-heating the finals or anything 
else in the RF line.  It sends a series of pulses to the transceiver allowing the 
peak power to be high enough to drive the amplifier to full power, but with the 
average power being low enough not to over-heat the final(s). 
 
The Ameritron ATP-100 Tuning Pulser  works with any solid state amateur 
transceiver or transmitter, and all other equipment using positive keying line 
voltages below 50 volts open circuit and 100 mA closed key current. It can also 
be used with negative voltage key lines (commonly found in older tube type 
radios) 

provided keying voltages are below 25 volts dc open circuit

.  

 
The Ameritron ATP-100 Tuning Pulser can also be used as a valuable 
troubleshooting and diagnostic aid.  The Ameritron ATP-100 Tuning Pulser  can 
be used in performance tests of QSK systems, wattmeters, and other equipment 
(see appendix). 
 
The Ameritron ATP-100 Tuning Pulser  uses a single 9-volt battery for power, 
with battery life dependent on operating and storage time.  To install the battery 
remove the cover by removing the two screws (one on each side) that secure it.  
A battery clip and holder, located on the left side of the enclosure, are provided 
for installing the 9-volt battery. 

Technical Description 

The Ameritron ATP-100 Tuning Pulser  uses a 555 timer driving an FET open 
drain switch. Two internal controls allow independent adjustment of both pulse 
repetition rate (PULSE RATE) and on-to-off time duty cycle (DUTY CYCLE). 
 
PULSE RATE is adjustable from about 17 to 40 pulses per second, while DUTY 
CYCLE is adjustable from about 10 to 90 percent. The keying time intervals are 
typically adjustable from 2- 40 mS "on" time, and a 24 to 60 mS pulse repetition 
time interval. 
Minimum duty cycle and the slowest PULSE RATE occur at the full 
counterclockwise settings of both internal controls.  The controls are designed to 
accept a small flat blade screwdriver. 

Summary of Contents for ATP-100

Page 1: ...cuit The Ameritron ATP 100 Tuning Pulser can also be used as a valuable troubleshooting and diagnostic aid The Ameritron ATP 100 Tuning Pulser can be used in performance tests of QSK systems wattmeter...

Page 2: ...ech processing increases the average power pushing the short term average power of normal speech to 30 or more During normal voice operation most equipment heat is generated by the quiescent current r...

Page 3: ...longer clockwise on time DUTY CYCLE settings If the pulse rate is too slow and or the duty cycle too slow peak power will not be reached All exciters have a delayed response to the leading edge of the...

Page 4: ...ck hazard and potential of accidentally keying the rig if the cabinet of the Ameritron ATP 100 Tuning Pulser or the operator comes in contact with grounded equipment WARNING Never use the ATP 100 with...

Page 5: ...ol to the rated drive power of the amplifier Place the amplifier into operate mode and depress the pulse tune button on the Ameritron ATP 100 Tuning Pulser Tune the amplifier for peak output power Not...

Page 6: ...ATP 100 Instruction Manual Tuning Pulser 6 Schematic...

Page 7: ...100 R1 Resistor 47K OHM 100 4470 R6 R7 Resistor Trimpot 500K OHM 133 5500 C1 C3 Capacitor 01uF 200 0416 C2 C4 Capacitor 1uF 200 0754 D1 D2 D3 Diode 1N4148 300 0140 D4 Diode 1N4007 300 0266 Q1 FET 2N70...

Page 8: ...measuring the single tone power with a barefoot exciter running off of a regulated supply and adjusting an oscilloscope to read near full screen with that carrier By activating the ATP 100 with the PU...

Reviews: