Illuminating Bulb Chamber (Fig. 13)
The bulb can easily and quickly be changed by opening the lower corner of the lensometer.
Fig. 13
Instrument’s Inclination Regulating Handle (Fig. 14)
This handle allows the operator to freely regulate the inclination of the instrument to work at a comfortable
posture. After screwing the handle outward, the operator can incline the device to a required angle according
to the state of the working table. Then lock the handle to secure the inclination.
Fig. 14
OPERATION INSTRUCTIONS
Preparation before Measurement
Place the instrument on the working table and adjust the inclination so that the operator is able to
make the measurement comfortably.
Connect with the mains.
Before measurement, visibility of the eyepiece shall be regulated in order to obtain accurate
results. While observing the black cross dividing line of the (8) eyepiece reticule, rotate the
eyepiece cover counterclockwise entirely. Then slowly turn the eyepiece cover clockwise until
the black cross becomes clearest. The eyepiece is now in focus.
Rotate the diopter measuring handwheel to regulate the scale interval to zero position. Observe
the green division line of (3) marker reticule in the view field of the eyepiece. The green dividing
line should now be clearest.