Gas Monitor 1ACO One-Channel Monitor
18
•
V
alarm
= (( 25 PPM / 100 PPM) * (1.320V – 0.380V)) + 0.380V
= 0.615V
RV5 would be adjusted so that 0.615V is measured at TP5.
+ -
MULTIMETER
DIGITAL
MEASURE VOLTAGE
AT SIG. TEST PT.
WITH RED (+) LEAD
CONNECT BLACK
( - ) LEAD TO GND.
PAD
TB1
TP7
RV6
RV4
RV5
TP1
TP2
TP3
TP4
TP5
TP6
JB8
JB10
1
R1
Figure 9: Trimmer adjustment and test point locations