4.1. Disassembly of the mask (recommended)
Unbutton or remove mask cuff from dome.
Caution!
Before autoclaving remove stopper from the inflation tube.
4.2. Cleaning of parts
Manual washing or automatic washing machine
Wash the parts in warm water using a detergent suitable for the mask
materials cf. paragraph 5. Rinse all parts thoroughly in clean water to
remove all detergent residues.
If surface cleaning and/or disinfecting of the outer surfaces of the
mask are carried out, make sure that the detergent and/or disinfec-
tant is compatible with the materials of the mask and be sure to
remove the detergent/disinfectant by rinsing with water. If residues
are allowed to dry on the mask, the lifetime of the materials may be
reduced.
An automatic washing machine with a programme designed for
washing anaesthesia breathing equipment may be used.
Disassembled mask with stopper
removed from the inflation tube.
Disassembled mask