background image

11

EN

WARNING

Avoid disturbing the mask during use.

7.8. Usage with Spontaneous Ventilation

The Ambu AuraOnce is suitable for spontaneously breathing patients when used with volatile agents or intravenous 
aesthesia on condition that anaesthesia is adequate to match the level of surgical stimulus and the cuff is not 
overinfl ated.
Coughing, breath-holding, or movement may occur if the level of anaesthesia is inadequate for maintenance. This 
may well occur following the introduction of an external stimulus such as surgery or turning the patient if the level 
of anaesthesia has been misjudged. Gently assist ventilation until breathing returns.

7.9. Usage with Positive Pressure Ventilation

Before using the Ambu AuraOnce with positive pressure ventilation (PPV), the operator should fi rst acquire
experience in its usage in spontaneously breathing patients. 
Choose a ventilatory pattern giving peak airway pressures less than 20 cmH

2

O and tidal volumes less than 8 ml/kg 

while the capnography is closely monitored.

In the event of leakage occurring during PPV, check for the following:
• light anaesthesia causing a degree of glottis closure
• inadequate neuromuscular block
• a reduction in lung compliance related to the surgical or diagnostic procedure
• displacement of the Ambu AuraOnce by head turning or traction.

After identifying the cause of the leakage, take appropriate corrective measures.

If leakage should occur around the cuff, 

do not simply add more air

. This will not necessarily improve the seal 

pressure and may even increase the leak by adding tension to the normally soft cuff, pushing it away from the 
larynx. Instead remove the mask and reinsert while providing that anaesthetic depth is adequate.

7.10. Critical observations during use 

Inadequate level of anaesthesia:

 The most likely problem following insertion is failure to maintain an adequate 

level of anaesthesia. Administer an additional bolus of induction agent and/or increase the concentration of volatile 
agent while gently assisting ventilation.

Incorrect positioning of the Ambu AuraOnce

 can be assessed by capnography, the observation of equal 

movements or by observation of changes in tidal volume, e.g. a reduction in expired tidal volume. If you suspect 
that the Ambu AuraOnce has been positioned incorrectly, remove and reinsert – and provide that anaesthetic depth 
is adequate. 

Unexpected regurgitation: 

Regurgitation may occur even in fasted patients. This may be caused by inadequate 

level of anaesthesia. One early sign of regurgitation is the appearance of fl uid travelling up the Ambu AuraOnce 
airway tube. The fi rst signs of regurgitation may be spontaneous breathing, coughing or breath-holding.

If regurgitation occurs, provided that oxygen saturation remains at acceptable levels, the Ambu AuraOnce should 
not be removed. This should be managed by putting the patient in a “head-down” position. Briefl y disconnect 
the anaesthetic circuit so that the gastric contents are not forced into the lungs. Check that anaesthetic depth is 
adequate and deepen anaesthesia intravenously, if appropriate.

Summary of Contents for AuraOnce

Page 1: ...Product Information Ambu AuraOnce Single Use Laryngal Mask Sterile For use by trained clinicians only...

Page 2: ...7 3 1 Placement technique 9 7 4 Insertion problems 9 7 5 Inflation 9 7 6 Connecting to the anesthetic system 10 7 7 Fixation 10 7 8 Usage with spontaneous ventilation 11 7 9 Usage with positive press...

Page 3: ...ery equipment Use of a nasogastric tube may make regurgitation likely because the tube may interfere with the function of the lower esophageal sphincter Do not attempt to clean and reuse the Ambu Aura...

Page 4: ...should only be used in patients who have been clinically evaluated by a clinician familiar with anesthesia as eligible for a laryngeal mask airway When the Ambu AuraOnce is used in profoundly unconsci...

Page 5: ...1 g Internal volume of ventilatory pathway 5 5 ml 8 ml 11 ml 15 ml 16 ml 21 ml 30 ml 38 ml Pressure drop Min interdental gap 15 mm 17 mm 19 mm 21 mm 25 mm 29 mm 31 mm 32 mm i Internal pathway 10 3 cm...

Page 6: ...shown in table 3 are maximum volumes Applying the stated maximum inflation volume may respond to a cuff pressure above the maximum of 60 cm H2 O It is recommended to continuously monitor the cuff pre...

Page 7: ...ot be any bulge nor any sign of leakage in the cuff pilot tubing or pilot balloon WARNING Do not use the Ambu AuraOnce if there are any bulges on the cuff or if there are any signs of leakage Table 5...

Page 8: ...ce are familiar with the warnings precautions indications and contraindications found in these Product Information The following points are extremely important Check for correct deflation and lubricat...

Page 9: ...nal or intravenous agents and initiate manual ventilation If you cannot open the patient s mouth sufficiently to insert the mask check that the patient is adequately anesthetized Ask an assistant to p...

Page 10: ...d to determine sufficient respiration Auscultate in the anterolateral neck region to check for abnormal sounds that might indicate mild laryngeal spasm or light anaesthesia The mask may leak slightly...

Page 11: ...o not simply add more air This will not necessarily improve the seal pressure and may even increase the leak by adding tension to the normally soft cuff pushing it away from the larynx Instead remove...

Page 12: ...ipment and the facility for rapid tracheal intubation are available Do not fully deflate the cuff until after its removal to avoid secretions entering the larynx and to prevent laryngospasm Alternativ...

Page 13: ...te that with the Ambu AuraOnce as with any form of airway management and anesthesia in pediatric patients where ventilation is insufficient desaturation is likely to occur faster because of the higher...

Page 14: ...14 EN c e 1 2 6 5 4 3 7 30 ml 60 cm H 2 O 50 7 0 kg 4 d A J I K H G D C B F E 2 6 5 4 3 1...

Page 15: ...15 EN g i h f...

Page 16: ...Catalog number LOT REF Copyright 2008 Ambu A S Denmark All rights reserved No part of this programme or the programme documentation may be reproduced in any form including photocopying without the pri...

Page 17: ...ermany Ambu GmbH In der Hub 5 D 61231 Bad Nauheim Tel 49 6032 92500 Fax 49 800 ambude www ambu de UK Ambu Ltd 8 Burrel Road St Ives Cambridgeshire PE27 3LE Tel 44 0 1480 498 403 Fax 44 0 1480 498 405...

Page 18: ...497 3600 01 11 2008 Version 1 0 1 Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 7225 2000 F 45 7225 2050 www ambu com 0086 LATEX FREE LATEX ATE Pat Pending...

Reviews: