EN
8.6
Remote Switches
Possible Problem
Possible Cause
Solution
The remote switches are
not working or not working
properly
Endoscope connector is not
properly connected to the
aBox™ Duodeno
Connect the endoscope con-
nector properly to the aBox™
Duodeno
Remote switches configura
-
tion changed
Return to standard configura
-
tion of the remote switches or
change the settings
Wrong remote switch
operated
Operate the correct remote
switch
aScope™ Duodeno defective
Connect a new aScope™
Duodeno
aBox™ Duodeno defective
Contact support
9.
Compliance And Licensing
The system conforms to the standard of IEC 60601–1 / ANSI/AAMI ES60601-1, IEC 60601–1–2 and
IEC 60601–2–18.
10.
Electromagnetic Compatibility
General information
Medical electrical equipment is subject to special
precautions with respect to EMC and must be in
-
stalled according to the instructions in the accom
-
panying documentation.
The manufacturer can only guarantee compliance
of the equipment if the accessory parts listed in
the accompanying documentation are used.
The device is intended for use exclusively by
trained medical personnel. This device can cause
radio interference or interference with the opera-
tion of other equipment in its close vicinity. It may
be necessary to take suitable corrective measures,
such as readjustment, a different system layout, or
shielding.
Special Instructions
Regulations applying to medical devices require us
to provide you with the following information. (See
the following pages of all tables.
Table 1 Recommended safety distances
Table 2 Electromagnetic compatibility 1
Table 3 Electromagnetic compatibility 2
Table 4 Electromagnetic transmission
Table 1 Recommended Safety Distance
Recommended safety distance between portable and mobile HF telecommunications systems
and the aScope™ Duodeno
The system is intended for operation in an electromagnetic environment in which HF interfe
-
rence is controlled. The user of the system can prevent electromagnetic interference by main
-
taining a minimum distance between portable and mobile HF telecommunications systems and
the system, depending on the output power of the communication unit, as specified below.
23