13
CALIBRATION OF TI-PREFORMS BLANK HOLDER
E N
2.4
Initial calibration of the Ti-Pre-
forms blank holder
The calibration process consists of two steps that
must be preformed one after the other.
1. Prepare the hybrid milling machine
2. Perform the calibration of the Ti-Preforms
blank holder
2.4.1 Preparation of the hybrid milling
machine
To ensure correct calibration of the Ti-Preforms
blank holder:
▷
Carry out auto-calibration: Follow the steps in
the Ceramill Motion 3 Service menu under
Cali-
bration.
▷
Perform the chuck cleaning: Follow the steps in
the Ceramill Motion 3 Service menu under
Clean
chuck.
NOTE:
Insufficient milling results when using in other
hybrid milling machines!
The blank holder is individually calibrated to the
respective hybrid milling machine.
▷
Afterwards, do not use the holder in other
hybrid milling machines.