AMALGAMATED INSTRUMENT HH4-PH Operation And Instruction Manual Download Page 4

2

Operation

The HH4-PH digital pH meter is operated by pressing the on/off button. The following

display sequence is executed during startup:

8.8.8.8

Display check

AIC

Company name

PH

X.X

Software version number

This sequence will take around eight seconds then the display will indicate the current

pH value. If a millivolt or temperature reading is required, press the “pH/mV/Temp” keypad to
obtain the required reading. The display may be cycled through “mV”, “temp” and “pH” by
successively pressing the keypad. The current display may be held (frozen) at any time simply
by pressing the “Cal/Hold” keypad.

Note: When the “Cal/Hold” keypad is pressed the display will indicate “Hold” and then

hold the current reading, if the “Cal/Hold” keypad is pressed again the display will indicate

LIUE

and the display will be released from hold.

The instrument features an automatic turn off function and will automatically power

down after a predetermined number of minutes (note: see chapter on setting up the instrument).
This feature is ideal for saving power, when the instrument is accidentally left turned on. The
automatic turn off function, may be disabled by selecting “0" minutes when setting up the
instrument.

2 .1

Read this first

Your new HH4-PH has been supplied factory configured (unless otherwise specified),

as follows:

Manual Temperature:

Set to 25

o

C

Temperature units:

Set to

o

C

Auto turn off:

Set to 10 minutes

Digital filter:

Set to 3 (range 0 to 8, 8=max)

Page 4 of 16

HH4PHMAN-1.0-2

Summary of Contents for HH4-PH

Page 1: ...HH4 PH Hand Held Digital pH mV Temperature Meter Operation and Instruction Manual...

Page 2: ...g up and programming functions 6 Step 1 Entering the setup mode 6 Step 2 Stepping through the settings 6 Step 3 Making changes to the settings 6 Programming Functions 7 Function Table 9 Calibration 10...

Page 3: ...ution of 1mV XXXX i e no decimal point places and the temperature range to 0 1 o XXX X i e one decimal point place Unless otherwise specified at the time of order your HH4 has been factory set to a st...

Page 4: ...ssed the display will indicate Hold and then hold the current reading if the Cal Hold keypad is pressed again the display will indicate LIUE and the display will be released from hold The instrument f...

Page 5: ...on Page 5 of 16 HH4PHMAN 1 0 2 Cal Hold keypad use also to enter setup mode and to step through the functions pH mV Temp keypad use also as down button when setting up On Off keypad use also as up but...

Page 6: ...f FUNC is followed by the first setup message OFSt As with all other messages the first display lets you know which parameter will be affected when changes are made Each time the Cal Hold button is pr...

Page 7: ...he HH4 PH is set to manual the manually set temperature will be displayed when the display is cycled to read temperature if the temperature probe is not connected When the HH4 PH is set to automatic t...

Page 8: ...bration values This function should only be used when calibration problems exist and it is necessary to clear the calibration memory See Calibration chapter for full details CAL E millivolt calibratio...

Page 9: ...anual temperature value FLtr Digital filter range 0 to 8 0to 8 8 most filtering APdt eSt auto turn off time 0 to 250 minutes Functions below are accessible via the calibration CAL mode only CAL1 Calib...

Page 10: ...f the pH reading has the same error at two points then the offset calibration OFSt should be used If the pH reading is correct around 7pH for electrodes with Eo 7pH and increases progressively above o...

Page 11: ...ble pH calibration 2 points Enter the Setup Mode in the usual way by pressing then releasing the Cal Hold button then pressing both the On OFF button and the pH mV Temp v button simultaneously within...

Page 12: ...al the first buffer value e g for a 7 00 pH buffer 4 Press Cal Hold to accept this new scaling value The message will be seen if the input has been accepted PH1 7 00 Stnd 2 Point calibration example F...

Page 13: ...through the functions by pressing the Cal Hold button until the display shows UC E Press the and v pushbuttons simultaneously the display will show CAL E CLR followed by FUNCEnd indicating that the ca...

Page 14: ...calibration and an attempt is made to calibrate then this message will be displayed The calibration will remain unchanged or display overrange The value attempting to be displayed is above 9999 or bel...

Page 15: ...Conversion Method Dual Slope ADC 4 samples per second Microprocessor MC1468HC05C85 CMOS Ambient Temperature 0 to 50 C Humidity 5 to 95 non condensing Display LCD 4 digit 12 7mm Power Supply 9V 216 st...

Page 16: ...manufacturer freight paid and if accepted for free repair will be returned to the customers address in Australia free of charge When returning the product for service or repair a full description of...

Reviews: