5
Step 4
Press and hold the blue “Temperature” button for 3
seconds whilst gently pivoting the probe from side to
side within the ear. This allows you to measure the
temperature of the entire ear canal to find the highest
temperature.
Replace the probe cover after use, to ensure an accurate measurement
and avoid cross contamination.
Only use Alvita probe covers.
Step 5
After 3 seconds, a long beep will be heard. This is when the measurement can be read.
A further two beeps will sound to indicate that the thermometer is ready to take another
measurement.
If the recorded temperature:
Is lower than 34°C (93.2°F)
The display will show “Lo”
(4 short beeps will sound).
Is between 34°C (93.2°F) and 37.5°C (99.5°F)
The display will show the temperature
1 long beep will sound).
Is higher than 37.5°C (99.5°F)
Fever Alarm:
The display will show the temperature flashing
(4 short beeps will sound).
Is higher than 42.2°C (108.0°F)
The display will show “Hi”
(4 short beeps will sound).
It is recommended that you take 3 measurements from the same ear replacing
the probe cover each time. If the 3 measurements differ, select the highest
temperature.