ALTONOVUS NanoRanger Quick Start Manual Download Page 2

 

 

 

USING THE FULLY ISOLATED USB INTERFACE 

NanoRanger’s micro

-USB interface can be used to export readings 

and to control the device via a PC (SCPI-format).  The USB interface 

cannot

 be used to power NanoRanger, as it is isolated from the main 

circuitry to prevent interference.  For help connecting NanoRanger, 

see the guidance in our website’s NanoRanger section.  For SCPI 

instructions, please download the full NR-01 Instruction Manual at: 

www.altonovus.com

 

 

USE IN CONJUNCTION WITH THE FULL INSTRUCTION MANUAL.  DOWNLOAD AT: 

WWW.ALTONOVUS.COM 

MODEL: NR-01: AUTO-RANGING DC AMMETER-
QUICK START GUIDE

 

NANORANGER - BASIC MENU FUNCTIONS 

This device complies with part 15 of the FCC Rules.  Operation is subject to the following two     
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any 
interference received, including interference that may cause undesired operation.  

NOTE: This equipment has been tested and found to comply with the limits for a Class B digital 
device, pursuant to part 15 of the FCC Rules.  These limits are designed to provide reasonable  
protection against harmful interference in a residential installation.  This equipment generates, uses 
and can radiate radio frequency energy and, if not installed and used in accordance with the     
instructions, may cause harmful interference to radio communications.  However, there is no     
guarantee that interference will not occur in a particular installation.  If this equipment does cause 
harmful interference to radio or television reception, which can be determined by turning the      
equipment off and on, the User is encouraged to try to correct the interference by one or more of 
the following measures:  

- Reorient or relocate the receiving antenna.  

- Increase the separation between the equipment and receiver.  

- Connect the equipment into an outlet on a circuit different that to 
which the receiver is connected.  

- Consult the dealer, or an experienced radio/TV technician for help.  

• 

Measure (as described on the previous page). 

• 

Range Maximum. 

• 

Range Minimum. 

• 

Measure Average. 

• 

Advanced. 

Press the Menu button to exit      
measuring and enter the Menu mode.  
This screen will appear.  Use the 
Navigate and Select buttons to move 
through the Menu.  The options are: 

RANGE MAXIMUM / MINIMUM - Use 
these options to set the 

upper and lower 

limits over which NanoRanger will        
Auto-Range during operation. 

If not constrained by the User, 

NanoRanger will exploit it’s full          

measurement range. 

The screen may show “UNDER LOAD” if 

a minimum range has been set, or if    
nothing is connected to NanoRanger. 

MEASURE AVERAGE - Use this function to calculate an average 
current over a time period, within a single range.  Having prevented 
auto-ranging (Set Range Max and Min), press Select to begin, and the 
first screen below will be shown.  Press Select again to end the meas-
uring period.  At the end of the period, the second screen shows: 

• 

The average current over the period 
(here 910.0uA). 

• 

The sample time, shown in multiples 
of 200mS. 

• 

The total charge for the period (here 
1.668uAh). 

• 

The highest (Sh) and lowest (Sl) 
samples taken in the period. 

ADVANCED MENU FUNCTIONS -   

In this element of the Menu, the User 
can configure the Auto-Powerdown 
timer and LCD back-Light. 