ALTONOVUS NanoRanger Quick Start Manual Download Page 1

 

 

MODEL: NR-01: AUTO-RANGING DC AMMETER-
QUICK START GUIDE (Rev 6.5 dated 14 Nov 20) 

ALTONOVUS LTD.  37 OAST HOUSE CRESCENT,         
FARNHAM, SURREY, GU9 0NP. UNITED KINGDOM

 

Do Not immerse in, 
or expose to liquid or 
dust.   

 

 

No User-serviceable     
components.  Do Not 
open unit, as it voids  
warranty and           
calibration. 

Do Not drop, expose to 
shock, or place heavy 
objects on unit.            
Do Not apply pressure 
to display.       

Avoid extremes of 
temperature.  Max 
accuracy in 20-25  
degree Celsius range. 

DEVICE

 USE AND CARE 

Only use a clean, 
dry cloth to clean 
unit. 

Do Not exceed  
maximum device 
load of 1.0A.  Higher 
current can cause 
device  failure. 

A - 

Positive

 Terminal 

B - Negative Terminal 

C - Display Screen 

D - On / Off Button 

E - Menu Navigation 

F - Select Button 

G - Isolated USB Interface 

H - Battery Compartment 

BEFORE USE 

If desired, remove the protective film from the front of 
the NanoRanger unit.  Peel carefully from below the 

Positive

 and Negative terminals.  NanoRanger will 

still function normally, if the film is left in place. 

INSERT 2 x AA BATTERIES 

1.  Remove battery compartment cover (H) applying  
gentle pressure to finger pad. 

2.  Ensure clean contacts on batteries. 

3.  Insert new batteries, observing polarity           
instructions in battery compartment. 

4.  Replace battery compartment cover. 

2 x AA alkaline batteries will give in excess of 150 
hours of continuous operation.  Treat all batteries in 

accordance with the manufacturer’s instructions.

 

USE IN CONJUNCTION WITH THE FULL INSTRUCTION MANUAL.  DOWNLOAD AT: 

WWW.ALTONOVUS.COM 

SCREEN GUIDE 

A

 - The current reading, displayed in mA, uA, nA, and pA.  In this example, 

0.309uA. 

B

 - The small, lower Range Bar shows which ranges are in use.  The   

upper, longer Range Coverage Bar shows movement across the ranges 
during the reading.  The box surrounding both, shows which range        
constraint has been input, via the menu.  In this example, none. 

C

 -  The Signal to Noise Ratio (SNR).  The SNR indicates how much 

noise, variation, and AC components are in the signal, followed by       
Maximum and Minimum samples during the reading. 

D

 - Maximum (Mx) and 

E

 - Minimum (Mn).  The Mx and Mn Readings 

shown are those that have occurred since this measurement started.    

Below that, the highest (Sh) and lowest (Sl) Samples are shown, from 
which the Readings have been derived. 

GETTING STARTED  -  TAKE A READING 

WARNING

.  Use 

NanoRanger with    
cables less than 3   
metres in length. 

 

STEP 1:  POWER ON 

After inserting batteries 
as shown, press the 
POWER button to start 
NanoRanger.   

 

Connect 

Positive

 

terminal (

RED

) to 

the higher voltage 
side of the current 
to be  measured. 

STEP 2:  CONNECT LEADS 

Connect Negative terminal (BLACK) 
to the lower voltage side of the cur-
rent to be measured.  Current 

must 

flow from 

Positive

 to Negative.    

Leads (not supplied) should be     
industry standard, uninsulated,    
4mm banana plug.  Leads of 0.3M 
will help minimise any noise. 

STEP 3:  TAKE MEASUREMENTS 

Sampling at a high rate to       
eliminate mains noise, Nano-  
Ranger takes about 3 readings 
per second.  It displays the current 
in mA, uA, nA, and pA.  In this 
example, 0.309uA. 

A Range Bar underneath the digits 
shows which ranges are in use.  
The Range Bar oscillates after 
each reading to show how quickly   
readings are occurring and to 
provide a visual confirmation that 
the device is functioning correctly. 

STEP 4:  POWER OFF 

After measurements 
are taken, press the 
POWER button 
again to return 
NanoRanger to 
Standby mode.   

Reviews: