The test-signals are scaled as follows:
GW-test signal
(Volt)
=
threshold
(phys. unit)x 10 V
measuring range (phys. unit)
Notice:
After adjusting the setpoints the potentiometers are to be sealed with anti-temper seal/
protective coating.
5 Starting up
If an adjustment (A-K-1K / A-D-1K) has been ordered in combination with the amplifier(-s) and/or
transducers(-s) it may be necessary for a slight fine adjustment nonetheless. This is due to possible various
environmental influences as well as to mounting etc.
If any visual damage or malfunctions are noticed, the measuring system must be switched off and marked
appropriately.
■
Mounting transducer and amplifier
■
Connect transducer to the amplifier
■
Connect multimeter to analogue output
■
Connect supply voltage – consider pin assignment
■
Allow the system about 30 min. to warm up
■
Check function and calibration of the system
Notice:
The allocation of transducer / amplifier is to be complied with. After replacing a transducer, the
calibration has to be checked.
It is to be noted that there is a slight dependence between zero-point adjustment and amplification (gain).