Altec AIR, LLC
P210WLP Series Air Dryer User’s Guide
Page 64 of 91
P017455 – Rev. C
9.11 Testing High Outlet Pressure Alarm
9.11.1
Make a note of the current
Outlet Pressure (
OUTP
) reading.
9.11.2
Loosen the retaining nut on the outlet pressure regulator.
9.11.3
Turn knob clockwise
until Outlet Pressure
(
OUTP
) reading climbs
over the high outlet
pressure threshold (7.5
PSI, 51.7 KPA default).
After one (1) minute, the
High Pressure Alarm
should appear on the display.
9.11.4
Turn Outlet Pressure
Regulator knob counter-
clockwise until Outlet Pressure
(
OUTP
) reading lowers to the
reading recorded in step 9.11.1
9.11.5
Tighten the retaining nut to
lock the setting in place