Self-Calibration of A and HZ lamps
After replacing A or Horizon lamps, you MUST use the
Self-Calibration
func-
tion to calibrate them.
1.
Go to
SETUP
>
ADMIN
>
RLAMP
.
2.
Use
▲
and
▼
to highlight either
A
or
HZ
.
3.
Press
ENTER
to access lamp replacement screen.
4.
Input lamp parameters “Remaining Hours, Remaining Cycles and Serial
number”.
5.
Use
▲
and
▼
to highlight “Self-Calibration” and press
ENTER
.
The self-calibration process takes 5 minutes to finish. The button on the screen
switches from "Calibrating…" to "Complete".
Press
ENTER
to return to the lamp replacement screen.