BYK-Gardner GmbH
System Configuration | 6
295 024 855 E 2106
spectro2guide Pro
71
If the differences are within your tolerances you can continue to measure with
your instrument. Otherwise calibrate your instrument on the white standard.
NOTICE
Every difference measurement is stored in the instrument. Transfer the measure-
ment data to “smart-chart” to document each test case.
6.3.6 Pro Variant
For
spectro2guide pro
special standards for jetness test and calibration are pro-
Exclusive Parts for Pro Variant [
NOTICE
Always use the LED flashlight included in delivery to check the cleanliness of your
black test and gray calibration standard – see “
Cleaning the Jetness Standards
6.3.6.1 Calibration Information
The jetness standards are fix assigned to your instrument. To check this go to
Con-
figuration
>
Calibration
>
Information
. Scroll down to
Jetness test standard
to view the details for the black test standard.