 Specifications

Human Care provides suitable training in safe 

operation of the lift and other accessories. 

Further information about slings, rail systems 

and other accessories can be found at www.

humancaregroup.com. 

o

Important! To ensure that the lift 

belt will not be damaged, it is 

important that the lift is balanced 

during moving, lifting and lowering 

operations. Make sure the lift bar is 

balanced, the hooks are set at the same 

length and the sling is suspended 

correctly. Never hold or pull the lift belt 

ENG

Lifting capacity, max.: 

55030

150 kg/330 lbs

55040 

220 kg/484 lbs

55050

300 kg/661 lbs

Dimensions: (WxHxL)

400x140x300 mm

Weight of lift: 

10.9 kg/24 lbs

Batteries:

2x12V, 2.3Ah

Emergency lowering: 

Electronic and mechanical

Class: 

Internally powered equipment

Type: 

IP rating lift:

IPX4 

IP rating hand control:

IPX5

Noise level:

<65dB (loaded and unloaded)

Lifting speed:

1.8 m/min / 3.0 m/min (unloaded only)

Lifting height:

2080 mm/ 81,9 in.

Start/stop: 

soft

Indicator:
Standby

flashing GREEN

Active

constant GREEN

Error

constant RED

Service

flashing YELLOW/GREEN, YELLOW/RED

Charging

constant YELLOW

Charger

Input 100-240V AC 50-60 Hz, 1.6A; Output 27.8V, 0.8A DC

IP rating charger:

IPX0

ETL:

Yes

This product is designed 

for repeated use. 

The telescopic bar is 

adjustable to three 

different widths: 40, 50 

and 60 cm. 

when moving a patient. It is also 

important that the lift belt is tensioned as 

it is wound in. 

o

Warning! Modification of the 

product is not permitted without 

the manufacturer’s consent! 

o

Important! If the lift is being used 

with a 2 point suspension method the lift 

should not be operated to its highest level 

(belt roll fully retracted) with weight 

attached to it.

Summary of Contents for 55030

Page 1: ...Altair Anv ndarmanual User manual Manuel de l utilisateur Gebruikshandleiding Benutzerhandbuch...

Page 2: ...raise and lower patients in a vertical direction and never diagonally or through angles The lift must always be positioned directly above the patient when lifting The patient must always be positioned...

Page 3: ...th Altair Before use always ensure that personnel who use the equipment have been given adequate tuition and training in use of the lift and its accessories the lift belt is not twisted or worn and th...

Page 4: ...hanical Class Internally powered equipment Type B IP rating lift IPX4 IP rating hand control IPX5 Noise level 65dB loaded and unloaded Lifting speed 1 8 m min 3 0 m min unloaded only Lifting height 20...

Page 5: ...www humancaregroup com Product description 1 Hand control 2 Lift belt 3 Telescopic bar 4 Leg hook 5 Secondary control panel 6 Emergency lowering button 7 Indicator 8 Emergency stop ENG 1 2 3 8 7 6 5...

Page 6: ...duct has been tested by an accredited test institute and complies with the necessary requirements in accordance with the Medical Devices Directive for Class 1 products MDD 93 42 EEC The lift is CE mar...

Page 7: ...falls If the lift belt is pulled out too quickly the safety catch will engage thus preventing the fall Slacken the lift belt to unlock the catch With the safety catch engaged the lever 10 can never be...

Page 8: ...run at two speeds If the Up or Down button is pressed for longer than 3 seconds the lift will run in the faster mode If the lift is loaded with more than approx 40 kg it will only run in the slower m...

Page 9: ...USB for communicating with the lift Secondary control panel The lift is equipped with a secondary control panel located on the lift which can be used as an alternative to the hand control It is equipp...

Page 10: ...level is Low The lift must be charged up immediately Charger The indicator on the lift flashes red when the batteries need recharging Plug the charger cable into the hand control or put the hand cont...

Page 11: ...patient Press the Down button on the hand control or control panel to lower the patient Pressing the button activates a soft start and releasing it causes a soft stop Lower the lift so that it stands...

Page 12: ...ve three adjustable positions To change position turn the hooks 1 4 turn while pushing or pulling the hooks to the desired position The hooks will automatically fall into place in the position and so...

Page 13: ...atient individually and in accordance with their needs Consult the user manual for the relevant sling to make sure the product is used correctly Human Care recommends using only Human Care slings with...

Page 14: ...or the Down button on the hand control or the control panel Constant yellow Charging in progress Interrupt charging Nothing happens when you press the Up button on the hand control or the control pane...

Page 15: ...he lift belts for damage or wear and tear Feed out enough of the lift belts to allow Care instructions for the lift and accessories their entire length to be checked Contact a retailer if the belt is...

Page 16: ...vel Compliance level Electro magnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor...

Page 17: ...4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80MHz to 2 5GHz 10 V m 800MHz to 2 5GHz 3 Vrms 3 V m 10V m Portable and mobile RF communications equipment should be used no closer to any part of the Human Care li...

Page 18: ...e lifts The Human Care lifts are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Human Care lifts can help prevent e...

Reviews: