![alpro Weekend-System II Installation Manual Download Page 5](http://html1.mh-extra.com/html/alpro/weekend-system-ii/weekend-system-ii_installation-manual_2908750005.webp)
5
2. Installation and Connection of the Weekend-Systems II
2.1. Form of delivery
• Weekend-System II
• Connection cables
• Installation manual (for technical personnel)
• Operating manual
• Device passport and handover protocol
2.2. Positioning on the treatment unit
The positioning of Weekend-System II at the treat-
ment unit should be stated in particular cases. The
risk of stumbling or obstacle must be excluded.
After installation the Weekend-System II must be
accessible easily. In particular, it must be paid atten-
tion that the tank can be filled easily.
2.3. Connection to the treatment unit
The Weekend-System II must only be connected by
an expert authorized by ALPRO in accordance with
the valid instructions for the drinking water decree.
In case of interference to the treatment unit, the
warranty conditions of the manufacturer have to be
followed.
For the connection of the air and water pipes please
use the pipes, T-pieces and couplings included in the
delivery contents.
a) Water supply
As a rule, the supply is connected with a flexible metal
tube (armoured tube included in delivery) directly
at the angle valve in the connection box of the treat-
ment unit. It can also be connected behind the lock
valve so that the Weekend-System II becomes
depressurised when switching off the treatment
unit!
If the water pressure in the incoming pipe is
higher than 6 bar, an appropriate water pressure
regulator must be assembled in front of the
Weekend-System II!