Page 30 of 33
Note: This connecter is not used in conjunction with the SADPmini logging software. See section 17.0
for cables used with the SADPmini logging software
19
SADPmini Protocol
Contact your supplier for the “SADPmini Protocol Manual”.
20
Instructions for Replacing Desiccant in the SADPminiEx
FOR USE IN SAFE AREA ONLY
1.
Fully extend the Desiccant head (1) of the SADPminiEx.
2.
Unscrew the Retaining Collar (2), turning it counter clockwise (anti-clockwise), which
will unfasten the desiccant filled head from the main body of the SADPminiEx.
3.
Gently pull out the Desiccant Head from the main body of SADPminiEx.
4.
Remove the Desiccant Head Base “O” Ring (3) and keep in safe place.
5.
Holding the Desiccant Head with the Desiccant Base Assembly (4) on top, fully
unscrew the Desiccant Head Base Assembly turning it counter clockwise (anti-
clockwise). The Desiccant Head base may require slight loosening by placing in a soft
jaw vice before this step is carried out. Do not fully unscrew with Head Base at the
bottom, as this will spill desiccant on floor.
ONLY USE SOFT JAWS. DO NOT USE
TOOLS OR VICE WITH HARD JAWS
.
6.
Dispose of used desiccant and re-fill with new up to a level of 3mm (5) from the bottom
of the screw thread.
7.
Gently insert the Desiccant Head Base Assembly in the desiccant and tighten.
USE
SOFT JAWS - DO NOT USE TOOLS OR VICE WITH HARD JAWS
.