Alpha MED Scientific MED-A64MD1A Product Manual Download Page 10

8

Amplification Settings

Acquisition-Bandwidth Settings

The MED64 Main Amplifier (64-CHANNEL MAIN AMPLIFIER) has a broad bandwidth range between 0.1Hz-10kHz 
which allows users to record several types of extracellular potentials. The amplifier has analog low-cut filter 
settings (high- pass filter) available at 0.1, 1, 10, and 100 Hz cutoffs and high cut filter settings (low-pass filter) 
available at 1, 2.5, 5, 7.5, and 10 kHz cutoffs, which can be selected in the Mobius software. Signals with 
frequencies higher than the value set for low-cut filter and lower than the value set for high-cut filter are 
acquired. 

Keep in mind that filtering distorts the signal. Thus, it is usually better to record the signal at a wide bandwidth 
and analyze the acquired signals using Mobius’s digital filter if necessary. The noise level for the MED64 System 
is very low due to its low-impedance electrodes. Usually a good signal-to-noise ratio can be achieved even with 
the high-cut filter (low-pass filter) of 10kHz. On the other hand, Setting the low cut filter (high-pass filter) with 
0.1 Hz is usually unnecessary except for recording of very slow signals such as from gastrointestinal smooth 
muscle. 

These recording bandwidth settings are recommended:

- fEPSPs:                                           1Hz -10 kHz 

                                                       (1Hz - 2.5 kHz is acceptable to achieve even better signal-to-noise ratio.)

- Neuron spikes (single unit):              100Hz - 10 kHz
- Cardimyocyte signals (FPD analysis): 1Hz - 1 kHz (To achieve the stability for the K

+

 release peak time)

Avoid recording with lower bandwidth than indicated above. 

Gain Setting

Raw signals are amplified by x10 with the 64-CHANNEL HEAD AMPLIFIER, then sent to the 64-CHANNEL MAIN 
AMPLIFIER. The Gain at the Main Amplifier is set in the Mobius software by changing the “Input Range” at the 
Mobius’ [Acquire MED64R2 Data] or [Acquire MED64R2 Data w/Stim] control panel.

The “Input Range” selected by the Mobius shows the maximum level of input signal. For example, if 5.0 mV is 
selected, any signals larger than 5.0 mV are NOT acquired. T

he smaller the maximum input signal levels 

are, the larger the gain is.

 The following chart shows the gain that can be achieved by each maximum input 

signal level (Input Range) as well as the input signal resolution. 

The following Input Range settings are usually recommended:

-fEPSPs:                                   5.0 mV
-Neuronal spikes (single unit):    2.3 mV
-Cardiomyocyte signals:             2.9 mV, Change  to 5.0 mV depending on the size of your signal. 

                                                (e.g. Select 5.0 mV when 1 Hz - 10 kHz is selected for recording bandwidth.)

Gain

Maximum input signal level

(Input Range set with Mobius)

Input signal resolution

x217

2.3 mV

0.07 

V

x172

2.9 mV

0.09 

V

x100

5.0 mV

0.15 

V

x40

12.5 mV

0.38 

V

x20

25 mV

0.76 

V

Summary of Contents for MED-A64MD1A

Page 1: ... Product Manual MED64 Main Amplifier 64 CHANNEL MAIN AMPLIFIER P N MED A64MD1A A low noise multi electrode array system for in vitro electrophysiology ...

Page 2: ...amages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document 2020 Alpha MED Scientific Inc All rights reserved Version 2 01 December 4...

Page 3: ...autions 2 MED64 Main Amplifier Components and Functions 4 MED64 Main Amplifier with Analog Output 6 MED64 Main Amplifier Set Up 7 Amplification Settings 8 Acquisition Bandwidth Settings 8 Gain Settings 8 Warranty 10 Specifications 10 ...

Page 4: ...t in fire or electric shock Excessive bending pulling or slicing of the cord should also be avoided Do not pull on the cord when you are disconnecting the power This could cause an electric shock Grasp the plug firmly when you disconnect the power supply Never touch the plug with wet hands as a serious electric shock could result Foreign Materials Ensure that no foreign objects e g needles coins s...

Page 5: ... symbols used on equipment and documentation Environmental Conditions Indoor use Altitude up to 2000 m Temperature 5 40 C Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Maintenance Clean the cabinet panel and controls with a soft cloth lightly moistened with mild d...

Page 6: ...HE1 then sent to the MED64 Main Amplifier where input signals are amplified further and digitized Stimulus current can be delivered via any of the 64 electrodes using the built in stimulator in the MED64 Head Amplifier Components and their Functions Front panel 1 POWER Turns amplifier ON and OFF Back panel 1 INPUT Signal Input Connects to the OUTPUT terminal on the 64 CHANNEL HEAD AMPLIFIER 2 DIO ...

Page 7: ...o the CONTROL INPUT on the 64 CHANNEL HEAD AMPLIFIER MED A64HE1 6 SIGNAL GND Ground terminal for signals 7 USB USB port for interface with a PC 8 DC INPUT DC power supply cord inserts here Accessories 1 DC Power supply unit 1unit 2 68pin cable 1 unit Connects the INPUT terminal to the OUTPUT on the 64 CHANNEL HEAD AMPLIFIER 3 USB cable 1unit For connecting the 64 CHANNEL MAIN AMPLIFIER USB termina...

Page 8: ...ment Connection for the MED A64MD1A to another equipment An extra MED Connector Cable and Accessories are available to help you to connect your MED64 Main Amplifier to another equipment Fig 1 Back panel for the MED A64MD1A Fig 2 Systen diagram for the MED64 Systen with another equipment Fig 3 Available accessories left and their channel assignments right ...

Page 9: ...dified incubator It is connected with a cable as long as 2m without noise or signal attenuation The MED64 System does NOT usually require a Faraday cage or vibration isolation table necessary for the conventional electrophysiology rigs It is recommended that the MED64 system be installed on a STABLE TABLE such as lab bench using aluminium foil except when the system is used in conjunction with oth...

Page 10: ...re recommended fEPSPs 1Hz 10 kHz 1Hz 2 5 kHz is acceptable to achieve even better signal to noise ratio Neuron spikes single unit 100Hz 10 kHz Cardimyocyte signals FPD analysis 1Hz 1 kHz To achieve the stability for the K release peak time Avoid recording with lower bandwidth than indicated above Gain Setting Raw signals are amplified by x10 with the 64 CHANNEL HEAD AMPLIFIER then sent to the 64 C...

Page 11: ...gnal level here Set high pass filter here Set low pass filter here Setting of acquisition bandwidth and Gain on the Mobius Acquire MED64R2 Data module The acquisition bandwidth is now set for 100 Hz 10 kHz and the Gain is x100 ...

Page 12: ...r controls improper maintenance modifications or service by anyone other than AMS or damage attributable to acts of God Specifications Amplifier Digitizer General Power supply unit Specifications may not be satisfied depending upon the type of computer or operating environments used Only for use in animal studies research Specifications and external appearance are subject to change without notice ...

Page 13: ...11 ...

Page 14: ...Product Manual MED A64MD1A ...

Page 15: ...entific Inc Saito Bio Incubator 209 7 7 15 Saito asagi Ibaraki Osaka 567 0085 Japan Phone 81 72 648 7973 FAX 81 72 648 7974 http www med64 com support med64 com Manufactured by Alpha MED Scientific Inc 2020 Alpha MED Scientific Inc ...

Reviews: