Chapter 2

 

 
Setting-up 

Supplying Power to the Alnor AVM410/A

IRFLOW

 TA410 

The Alnor AVM410/A

IRFLOW

 TA410 is powered with four size AA 

batteries. 

Installing the Batteries 

Insert four AA batteries as indicated by the diagram located on the 
inside of the battery compartment. The Alnor AVM410/A

IRFLOW

 

TA410 is designed to operate with either alkaline or NiMH rechargeable 
batteries, although it will not recharge NiMH batteries. Battery life will 
be shorter if NiMH batteries are used. Carbon-zinc batteries are not 
recommended because of the danger of battery acid leakage. 

Using the Telescoping Probe 

The telescoping probe contains the velocity, temperature, and humidity 
sensors. When using the probe, make sure the sensor window is fully 
exposed and the orientation dimple is facing upstream.  

NOTE

:  For temperature measurements, make sure that at least 3 inches 

(7.5 cm) of the probe is in the flow to allow the temperature sensor to be in 
the air stream.

 

Extending the Probe 

To extend the probe, hold the handle in one hand while pulling on the 
probe tip with the other hand. Do 

not

 hold the cable while extending the 

probe as this prevents the probe from extending. 

Retracting the Probe 

To retract the probe, hold the handle in one hand while gently pushing 
on the probe tip with the other hand. If you feel the probe antenna 
binding, pull gently on the probe cable until the smallest antenna section 
is retracted. Collapse the rest of the antenna by pressing the probe tip. 

 

Summary of Contents for AVM410

Page 1: ...Ventilation Testing Air Velocity Meter Alnor Model AVM410 AIRFLOW TM Model TA410 Operation and Service Manual...

Page 2: ...nd is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE TO THE EXTENT PERMITTED BY LAW THE EXCLUSIVE REMEDY OF...

Page 3: ...er to the Alnor AVM410 AIRFLOW TA410 3 Installing the Batteries 3 Using The Telescoping Probe 3 Extending The Probe 3 Retracting The Probe 3 CHAPTER 3 OPERATION 5 Keypad Functions 5 CHAPTER 4 MAINTENA...

Page 4: ......

Page 5: ...tification Carefully unpack the instrument and accessories from the shipping container Check the individual parts against the list of components below If anything is missing or damaged notify TSI imme...

Page 6: ......

Page 7: ...e velocity temperature and humidity sensors When using the probe make sure the sensor window is fully exposed and the orientation dimple is facing upstream NOTE For temperature measurements make sure...

Page 8: ......

Page 9: ...tware Revision and Last Date Calibrated ft min m s Key Pressing this key changes the display to read air velocity C F Key Pressing this key changes the display to read temperature Changing Units To ch...

Page 10: ......

Page 11: ...ernational UK TSI Incorporated TSI Instruments Ltd Alnor Products Lancaster Road 500 Cardigan Road Cressex Business Park Shoreview MN 55126 3996 High Wycombe Tel 800 874 2811 Buckinghamshire 651 490 2...

Page 12: ......

Page 13: ...auses Corrective Action No Display Unit not turned on Switch unit on Low or dead batteries Replace batteries or plug in AC adaptor Dirty battery contacts Clean the battery contacts Velocity reading fl...

Page 14: ......

Page 15: ...ative humidity up to 80 RH non condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II External Meter Dimensions 4 2 in 7 2 in 1 5 in 10 7 cm 18 3 cm 3 8 cm Meter W...

Page 16: ...kinghamshire HP12 3QP England Tel 44 0 149 4 459200 Fax 44 0 149 4 459700 E mail info airflowinstruments co uk Web www airflowinstruments co uk France Tel 33 491 95 21 90 E mail tsifrance tsi com Webs...

Reviews: