4.5. Monitoring limit values (P3 und P4)
Force gauges series FMI-B30 and FMI-S30 and higer can compare the
actual measured value with individually set upper or lower limit valu-
es. The surpassing of the thresholds is indicated on the display and
signal outputs are switched accordingly. To utilize the output signals
a data cable for digital I/Os (part no.: FMI-934SO) is needed
28 chapter 5.2.3 Data cable for digital I/O’s)
. The limit function re-
places the overload function of the instrument (provided the set limit
value does not exceed the admissible overload).
4.5.1. Working with limit values
If a limit value is set (at least one of the limit values is uneven zero)
the limit function is activated and the display shows the Limit symbol.
As soon as the measured value is smaller than the lower limit value
a downward pointing arrow is shown on the right side of the display.
Accordingly an upward pointing arrow is shown if the measured value
surpasses the upper limit value. No arrow appears if the measured
values are within the limits.
A reset of the instrument sets limit values to zero.
4.5.2. Setting limit values
Select the menu P3 to set the upper limit values and P4 to set the
lower limit values.
Pressing the O-key you select the digit. There are six digits (a prefix
and five numbers). The active digit flashes. By pressing the I-key you
select the desired number. Any pressing of the key increments by one.
By pressing the S-key you save the selected values and return to the
upper menue level.
To set the lower limit value, select menue P4 and proceed as descri-
bed above.
activate
limit menue
select digit
select value
Limit
> <
> <
> <
> <
> <
> <
prefix
Limit
Limit
Digit 1
Limit
Limit
Digit 2
Limit
Limit
Digit 3
Limit
....
Digit 4
Limit
...
Digit 5
Limit
Limit
Setting the upper limit
F
MAX
50
0
50
OUT
Hirose
HI
LO
F
MIN
Working with limit values