5. Maintenance

5.1. Cleaning and disinfection

Caution: The dirt on the lens must be cleaned immediately after using. It will be hard to clean when the dirt
air-dried.
5.1.1 Cleaning the surface of equipment
The outer surface of the equipment can be cleaned with wet cloth. The remaining stains can be cleaned off
with the mixture of 50% distilled water. Do not wipe with any corrosive detergent lest that the surface should
be damaged.
5.1.2 Cleaning the surface of optical lens
To prevent the dust stained on the lens, never expose the optical lens to air without objective, eyepiece tube
and eyepiece. Please use the dust cover after using.
Cleaning the surface of optical lens: To clean the dirt on lens, such like bloodstain, please use the special paper
or absorbent cotton with little bit distilled water and wash, the remaining stains can be cleaned off with the
mixture of 50% Ethanol and 50% aether. If there is dust stained on the lens, blow them with a blow ball or
brush them with a dust pen. Do not wipe with any corrosive detergent lest that the lens should be damaged.

5.2 After-sales service

If any operator requests more information such as circuit diagrams, parts list and product descriptions, for
repairs carried out by qualified technical personnel, please contact us.

6. Trouble-shooting guide

In case there is any trouble, please first refer to the trouble-shooting guide. If it still can not work, please
contact the authorized distributor.

Trouble

Possible reason

Remedy

No working

Power adapter is not connected

Connect the power adapter

No switch on the main power

Switch on the main power

Main power is broken-down

Contact a local electrician

Illumination
does not light

The LED lamp is not connected correctly Connect again

No switch on the main power

Switch on the main power

Main power is broken-down

Contact a local electrician

Electric parts are broken-down

Contact the authorized distributor

The LED bulb is burnt

Replace the LED bulb

Temperature
rise suddenly

Something covered the heat
elimination holes

Take off the cover and clean the heat
elimination holes

Summary of Contents for YDJ-II

Page 1: ...1 YDJ II LED COLPOSCOPE OPERATION INSTRUCTION...

Page 2: ...them The environment for use a Transport and Storing Ambient Temperature Range 40 55 Relative Humidity Range 10 80 Barometric Pressure Range 500 hPa 1060 hPa b Operation Ambient Temperature Range 5 40...

Page 3: ...igh temperature Use it in the room and simultaneously be careful to keep it clean and dry 3 Check that all the wires are correctly and firmly connected before using 4 Please pay attention to all the r...

Page 4: ...camera adaptor demonstrator photo adaptor 3 6 Assembly of the binocular head 3 7 Connection of the LED lamp plug 3 8 Connection of the power adapter 4 Use of equipment P10 4 1 Necessary working condi...

Page 5: ...in the vagina which is impossibly viewed with naked eyes can be clearly seen with the instrument Therefore the correctness of a diagnosis is enhanced significantly Examination can be performed with v...

Page 6: ...6...

Page 7: ...7...

Page 8: ...8 Fig 4 Power adapter...

Page 9: ...he monitor in order to transfer the image to the monitor for better observation and image memory and processing 2 13 Beamsplitter It is used to divide the light 50 50 It is provided with connections a...

Page 10: ...package take off the threaded cover 33 Fig 2 insert it into the hole of arm then tighten this threaded cover 33 Fig 2 3 5 Assembly of the beamsplitter 45 tilted connecting component CCD camera adapto...

Page 11: ...he binocular heads with eyepieces from the microscope and observe the distant object just like using a telescope Then rotate the diopter adjustment hoop of one eyepiece until the image clear If necess...

Page 12: ...Turn on the power switch 36 Fig 3 Move the arm up and down to a right working position Adjust the illumination Select the filter which will be used Move the colposcope into operation area then adjust...

Page 13: ...them with a blow ball or brush them with a dust pen Do not wipe with any corrosive detergent lest that the lens should be damaged 5 2 After sales service If any operator requests more information suc...

Page 14: ...f 280mm Focal length of objective f 300mm or f 280mm Fine focusing 10mm Binocular 45 inclined Adjustable range for pupil distance 55 75mm Linear field 15 90mm f 300mm 13 4 84mm f 280mm Eyepiece magni...

Reviews: