Alltion (Guangxi) Instrument Co., Ltd.
AM-2000
User’s Manual | Safety Caution
3
Storage Environment
Ambient temperature range: -40
℃~
+55
℃
Relative humidity range: 10% ~ 60% RH
Atmospheric pressure range: 500hPa ~ 1060hPa
Well ventilated indoor environment free of corrosive gas or other harmful substances.
Safety Characteristics
Adaptor: Input: DC 12V / 3A
Classification according to the degree of protection of the liquid: IPX0;
Standard IEC60601-1 Type II Equipment.
Safety Requirements on Installation and Use
Safety Requirements
This instrument can be used only for the purposes described in the User's Manual.
Only trained and instructed personnel are allowed to use this instrument. The customer
or the organization operating the equipment has the responsibility to train and guide all
personnel using the equipment.
Before starting the instrument, please completely comprehend the User's Manual,
including the User's Manual for accessories and other system components.
Keep the User's Manual in order to facilitate the operator to read at any time.
Please observe all symbols and labels of the instrument!
The modification and repair of this instrument can only be performed by the ALLTION
service personnel or others authorized by ALLTION.
Do not place any container filled with liquid over the instrument. Confirm that no liquid
can penetrate into the instrument.
No modification of this equipment unless authorized by the manufacturer.
Although the equipment conforms to the intent of the standard IEC 60601-1-2 in relation to
electromagnetic compatibility, electrical equipment may produce interference. If interference
is suspected, move equipment away from sensitive device or contact us.
Do not store or use the instrument in a damp room. Do not expose the instrument in the
place with splashing, dropping or water mist.
When the instrument is generating smoke, electric spark or a strange noise, please
immediately cut off power supply of the instrument. Do not use this instrument until it has
been repaired by our service agent.
Please note that local regulations take precedence over the requirements of the above
mentioned criterions. If you have any enquiry, please contact the local ALLTION dealer.
Requirements on Installation
The installation of the product will be completed by our service representative or by
professional personnel authorized by us. Please make sure that the following operational
requirements have been met:
All the safety-related mechanical connections (please find the Manual for details) are
properly connected and all screws have been tightened.
Summary of Contents for AM-2000
Page 9: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Safety Caution 5 1 4 3 2 3 5...
Page 11: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Assembly 7 2 1 3 4 5 7 8 6...
Page 14: ...Alltion Guangxi Instrument Co Ltd 10 Product Assembly AM 2000 User s Manual 17 19 20 18...
Page 34: ...Alltion Guangxi Instrument Co Ltd 30 Product Functions AM 2000 User s Manual 1 2 3 4 5 6 7 7 7...
Page 36: ...Alltion Guangxi Instrument Co Ltd 32 Product Functions AM 2000 User s Manual 5 7 3 1 2 4 6 7...