Alltion (Guangxi) Instrument Co., Ltd.

2

Safety Caution

|

AM-2000 User’s Manual

Upward

Keep dry

Handle With Care

2

The number of stacked layers must not exceed 2 layers

Ambient temperature range: -40

~55

Relative humidity range: 10%~80%

1060hPa

500hPa

Atmospheric pressure range: 500 hPa~1060 hPa

Structure and Composition

AM-2000 Surgical Microscope consists of support bracket (floor-mounted mobile bracket
(standard)/optional floor-mounted bracket, ceiling bracket, high wall-mounted bracket and low
wall-mounted bracket (optional)), cross arm bracket (including electric installation) and
microscope head (including visual optical component, microscope body, objective lens and
illuminating system).

Range of Application

The AM-2000 surgical microscope is a manual microscope for surgical microsurgery
amplification, illumination, etc. It can assist you in microsurgery and fine examination in
departments such as otolaryngology and dentistry.

Contraindication

Patients with light allergy

Shall not be used in ophthalmology

Do not directly radiate it to the human eyes

Service Life

8 years.

Working Environment

Ambient temperature range: 5

~ 40

Relative humidity range: 10% RH ~ 80% RH

Atmospheric pressure range: 700hPa ~ 1060hPa

Summary of Contents for AM-2000

Page 1: ...AM 2000 Surgical Microscope User s Manual Please read the instruction carefully before using it PA1805 UM01 EN Version A4 2019 12 25 Alltion Guangxi Instrument Co Ltd...

Page 2: ...k Wuzhou Guangxi China E mail sales alltion com or sales alltion microscope com Tel 86 774 2836101 Fax 86 774 2836192 Postal Code 543100 Website http www alltion com After Sale Service Provider Alltio...

Page 3: ...of Fixed Floor Stand 13 Installation of Ceiling Mount 14 Installation of Low Position Wall Mount 15 High Position Wall Mount 16 Installation of Cross Arm System 17 Upright Installation of Cross Arm 17...

Page 4: ...s declaration electromagnetic emission 37 Guidance and manufacturer s declaration electromagnetic immunity for all EQUIPMENT and SYSTEMS 38 Guidance and Statement from Manufacturer Electromagnetic Imm...

Page 5: ...user of the instrument or providing the useful information to the user Information Symbols The item symbols used in this User s Manual are defined as follows The content of the upper and lower items i...

Page 6: ...ed bracket optional cross arm bracket including electric installation and microscope head including visual optical component microscope body objective lens and illuminating system Range of Application...

Page 7: ...e personnel or others authorized by ALLTION Do not place any container filled with liquid over the instrument Confirm that no liquid can penetrate into the instrument No modification of this equipment...

Page 8: ...ient Do not disassemble or assemble the supporting components during use to prevent the Support system from being out of balance damaging the components or hurting the patient This surgical microscope...

Page 9: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Safety Caution 5 1 4 3 2 3 5...

Page 10: ...straight or 45 bevel 1 3 Eyepiece 12 5X or 10X 2 4 Magnification pod five range zooming 1 5 Control handle 1 6 Objective lens F250 objective lens optional F300 objective lens please find the Optional...

Page 11: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Assembly 7 2 1 3 4 5 7 8 6...

Page 12: ...e binocular head Straight tube 2 45 binocular head 45 inclined 3 200 300 Variable Objective Lens Continuous zoom from 200mm to 300mm 4 F200 objective lens F200 5 F250 objective lens F250 6 F300 object...

Page 13: ...evice 13 Camera adapter Interfaces with Sony Canon Nikon cameras optional 14 2 D Rotation Binocular Assistant Scope Connector 15 3 D Rotation Binocular Assistant Scope Connector 16 ALL CAM2 Full Funct...

Page 14: ...Alltion Guangxi Instrument Co Ltd 10 Product Assembly AM 2000 User s Manual 17 19 20 18...

Page 15: ...not provided if any component is not provided please contact the local dealer in time Please check the product if there s any damage especially the optical components if any please contact the local d...

Page 16: ...put it aside Put base 1 on the horizontal ground and lock four casters Fit the plastic cover 2 onto the base 1 and tighten the compression ring 3 Insert the upright post 4 in the installation hole of...

Page 17: ...d with 14 driller the hole depth is 75mm and the hole position is as shown on the right diagram 170 170 14 If the position and dimension of the installation hole are incorrect the stand can t be insta...

Page 18: ...ram When drilling please note the direction of power cord connection The power connector is along same direction with the pre installed socket on the ceiling Lock the ceiling mount assembly 2 to the c...

Page 19: ...t shall be smooth enough otherwise it will cause the product to be tilted after installation Drill 6 holes in the wall with 14 driller the hole depth is 75mm The position and dimension of the hole is...

Page 20: ...shall be smooth enough otherwise it will cause the product to be tilted after installation Drill 6 holes in the wall with 14 driller the hole depth is 75mm The position and dimension of the hole is as...

Page 21: ...ing bolts 2 on the cross arm remove the plastic cover 1 and put it aside Lift the cross arm over the pole cross the power cable 3 from the mounting hole of cross arm 4 as shown by the bellow arrow Ins...

Page 22: ...the column insert the decorative ring 2 into the column insert the power plug 3 on the column into the power connector 4 Tighten with the nut 5 on the power plug Insert the connecting shaft on the fi...

Page 23: ...ctive lens 300mm objective lens 350mm objective lens 400mm objective lens and 200 300mm objective lens are the same Mount the large objective 1 into the microscope mount with two M4 screws 3 Install t...

Page 24: ...ion Align the binocular tube 1 at the positioning pin and install it in the bayonet of microscope body 3 Tighten the locking screw 2 Before loosening the grip please make sure the binocular head barre...

Page 25: ...Product Assembly 21 Wiring Connect the attached power wire according to the following picture If the specification of power cord plug does not match the local socket standard please contact your loca...

Page 26: ...ob of each joint is valid judge whether the locking function is normal Turn on the power switch observe whether the indicator of the power switch turns on whether the light is illuminated from the mic...

Page 27: ...on method is the same as the Installation of Binocular Head Barrel see Page 20 Installation of Objective Lens For objective lens of other specifications the installation methods are the same as Large...

Page 28: ...ctions Product Components 1 Main body of microscope 2 180 binocular and eyepiece 3 Objective lens 4 120 hanger bracket 5 Cross arm bracket 6 Mobile floor stand 7 Fixed Stand Mount 8 Low position wall...

Page 29: ...interface For 12V 1 0A DC output 3 Lighting shift knob For switching different illumination means large spot without color filter means medium spot without color filter means small spot without color...

Page 30: ...provide diopter compensation at 6D and 6D Setting the diopter adjustment at 0D if the operator wear glasses Rotating the diopter adjustment to the best position till you see the most clear view if the...

Page 31: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Functions 27 Objective lens 1 Focusing knob It is used for manually setting the image definition focusing working distance 1...

Page 32: ...rotation of the main lens of the microscope as indicated by the arrow on the left 2 Friction adjustment knob for tilting motion left right of the main lens of the microscope Used to adjust the fricti...

Page 33: ...ll accessories adjust the balance of the balance arm with the knob the adjusting identification as the one in the left picture For rotating the knob easily the balanced arm shall be in the horizontal...

Page 34: ...Alltion Guangxi Instrument Co Ltd 30 Product Functions AM 2000 User s Manual 1 2 3 4 5 6 7 7 7...

Page 35: ...sories adjust the balance of the balance arm with the knob the adjusting identification as the one in the left picture For rotating the knob easily the balanced arm shall be in the horizontal position...

Page 36: ...Alltion Guangxi Instrument Co Ltd 32 Product Functions AM 2000 User s Manual 5 7 3 1 2 4 6 7...

Page 37: ...cope Check before use There must be no patient when checking the microscope To terminate the operation of the equipment please turn off the power switch or disconnect from the network power supply Che...

Page 38: ...ched Adjust the surgical microscope to the maximum magnification and move the surgical microscope to the position that has clearest image Clearer image can be obtained by fine adjustment of the focus...

Page 39: ...Instrument Co Ltd AM 2000 User s Manual Performance Parameters 35 Performance Parameters Basic dimension The basic dimensions of the whole machine and the motion range of the joint are shown in the f...

Page 40: ...shift Orange filter green filter large spot without filter medium spot without filter small spot without filter Optical parameters Large objective lens F200 F250 F300 F350 F400 Variable magnific ation...

Page 41: ...decreasingly in electromagnetic compatibility transmission and immunity performance to use or replace with the accessories with non matched design Do not replace the components without authorization...

Page 42: ...V differential mode 1 kV differential mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply in...

Page 43: ...smitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m b Field strengths from fixed RF transmitters as determined by an electromagnetic s...

Page 44: ...ration distance according to frequency of transmitterm 150kHz 80MHz d 3 5 V1 p 80MHz 800MHz d 3 5 E1 p 800MHz 2 5GHz d 7 E1 p 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12...

Page 45: ...t free box Do not use any chemical cleaning agent corrosive solvent or detergent with scratching effect which will damage the surface of the optical device The stain blood stains etc on the surface of...

Page 46: ...ntermittently during use The cooling window and the air inlet are covered or blocked by external object Remove the foreign object and clean the cooling window Failure of cooling fan Contact the local...