WARNING !
•
DO NOT USE THIS COMPRESSOR IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
•
NEVER TRY TO PUT OXYGEN INTO THE COMPRESSOR. THE COMPONENTS ARE NOT APPROVED FOR
OXYGEN SERVICE.
•
REFER ALL REPAIRS TO QUALIFIED SERVICE PERSONNEL.
CAUTION !
•
DO NOT
use the PCS-414 with blenders, respirators or other respiratory equipment that requires dry air without
optional air cooler.
•
When making attachments to the outlet
DO NOT
use wrenches or excessive force. This can cause damage to
internal fittings and tubing.
•
DO NOT
lubricate any part of the compressor with oil, grease or petroleum products. If these substances come
in contact with the diaphragm, they can adversely affect the service life of the compressor.
•
DO NOT
substitute parts. Such substitutions can cause premature product failure and may void warranty.
II. UNPACKING INSTRUCTIONS
A.
Unpack and inspect for shipping damage.
B.
Before operating the compressor, remove the (4) shipping bolts from the bottom of the unit. This will
allow the compressor to hang freely inside the cabinet. Retain the (4) shipping bolts for future shipping of
the PCS-414.
C.
Uncoil the power cord and plug the appliance connector into the back of the PCS-414 compressor.
You can now plug it into a receptacle marked “Hospital Grade”.
D.
Turn the power switch on and verify that the unit functions properly.