WARNING !

 

 

• 

DO NOT USE THIS COMPRESSOR IN THE PRESENCE OF FLAMMABLE ANESTHETICS.

 

 

• 

NEVER TRY TO PUT OXYGEN INTO THE COMPRESSOR.  THE COMPONENTS ARE NOT APPROVED FOR 

OXYGEN SERVICE.

 

 

• 

REFER ALL REPAIRS TO QUALIFIED SERVICE PERSONNEL.

 

 

 
 
 
 
 

CAUTION !

 

 

• 

DO NOT 

use the PCS-414 with blenders, respirators or other respiratory equipment that requires dry air without 

optional air cooler.

 

 

• 

When making attachments to the outlet 

DO NOT 

use wrenches or excessive force. This can cause damage to 

internal fittings and tubing.

 

 

• 

DO NOT 

lubricate any part of the compressor with oil, grease or petroleum products.  If these substances come 

in contact with the diaphragm, they can adversely affect the service life of the compressor.

 

 

• 

DO NOT 

substitute parts.  Such substitutions can cause premature product failure and may void warranty.

 

 
 
 

II. UNPACKING INSTRUCTIONS 
 

 
A. 

Unpack and inspect for shipping damage. 

 

B. 

Before operating the compressor, remove the (4) shipping bolts from the bottom of the unit. This will 
allow the compressor to hang freely inside the cabinet. Retain the (4) shipping bolts for future shipping of 
the PCS-414. 

 

C. 

Uncoil the power cord and plug the appliance connector into the back of the PCS-414 compressor.     

You can now plug it into a receptacle marked “Hospital Grade”. 

 

D. 

Turn the power switch on and verify that the unit functions properly. 

 
 
 
 
 
 
 
 
 
 
 
 

 

4

 