Specifications
72
Goldeye G/CL Technical Manual V4.2.0
Operating conditions
Absolute QE
Parameter
Values
Operating temperature
-20 °C to +55 °C (case)
Storage temperature
-30 °C to +70 °C (ambient)
Temperature setpoints
1
•
+25 °C (default and calibrated)
•
+20 °C, +35 °C, +50 °C (other setpoints) or user-configurable
Relative humidity
10% to 95%, non-condensing
Power requirements (DC)
10.8 V to 30.0 V
Max. power consumption
10 W (12 VDC)
Typical power consumption
without cooling
4.5 W (12 VDC)
1
Even though the first temperature setpoint is pre-configured to +20 °C, it is possible to set it to a lower
temperature. However, if the sensor temperature is lower than the ambient temperature, especially in humid
environments, condensation may occur.
Table 12: Goldeye CL-008 TEC1 operating conditions
Figure 10: Goldeye CL-008 TEC1 absolute QE
0
10
20
30
40
50
60
70
80
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
Qu
a
n
tu
m
E
fficiency
[%]
Wavelength
[nm]
Goldeye
G/CL
Ͳ
008
Absolute
QE