Command reference
R&S
®
ZND
929
User Manual 1173.9557.02 ─ 61
[SENSe:]CORRection:CKIT:LDELete
<CalKitName>, <KitLabel>
Deletes an imported or user-defined cal kit which is identified by its cal kit name and
label.
Note:
It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel>
String parameter containing the label of an imported or user-
defined calibration kit available on the analyzer.
Example:
See
[SENSe:]CORRection:CKIT:LLABel
Usage:
Setting only
[SENSe:]CORRection:CKIT:LLABel
<CalKitName>, <KitLabel>[, <NewKitLabel>]
Assigns a calibration kit label to an imported or user-defined calibration kit or renames
an existing calibration kit label.
Parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel>
String parameter containing the current calibration kit label.
<NewKitLabel>
String parameter containing the new calibration kit label.
SCPI command reference